Anaesthetic Respiratory Equipment Conical Connector Tester EN ISO 5356-1:2015
|
Detailed Product Description
|
EN ISO 5356-1:2015 Anaesthetic and respiratory equipment — Conical
connectors Test equipment Check compliance by engaging the cone or socket in the appropriate
plug or ring test gauge according to Figure A.1 and Table A.1, by
applying an axial force of (35 ± 3,5) N for 8,5 mm, 11,5 mm, and 15
mm cones or sockets and (50 ± 5) N for 22 mm and 30 mm cones
or sockets while simultaneously rotating the cone or socket (20 ±
5)°. Its leading edge shall lie between the minimum and maximum
diameter steps of the gauge. The cones or sockets and gauges
shall be maintained at a temperature of (20 ± 3) °C during the
test. NOTE: Because cones or sockets made from plastics materials (for example polyamide, polyacetal, polycarbonate, polysulfone) can vary greatly in their physical characteristics, it is not considered practicable to specify dimensions A, B, and ratio F; for this reason, gauging requirements have been included. It is also considered impracticable to generalize on matters such as cold flow and thermal instability, as well as possible changes in physical characteristics, contact with solvents, etc. We have the following cones and sockets in our portfolio.Feel
free to send us an RFQ to sales@kingpo.hk
|
| Product Tags: Conical Connector Tester Respiratory Equipment Connector Tester EN ISO 5356 Connector Tester |
Related Products
|
Oxygen Environment IEC 60601-1 Spark Ignition Tester For Me Equipment / Systems |
|
Anaesthetic Respiratory Equipment Conical Connector Tester EN ISO 5356-1:2015 |
|
IEC 60335-1-Annex B-B.20.1 2070kPa Manual Pressure Tester 20 30 50 100ml |
|
IEC 60335-2-15:22.106 Electrical Safety Tester For Kettle Plug / Pull Out |
|
IEC 60335-2-84 Toilet Seat Endurance Test Apparatus Applying 890N |
|
10A 3 Station IEC60335-1 Flexing Test Apparatus |
Email to this supplier