Peltier Water Bath Thermoelectric Cooling Bath For Diffusion Gas
Standard Features
Fast cooling and heating with precise temperature
Easy to operate
Environmentally friendly and refrigerant-free
Programmable Digital Temperature Controller
System Error Monitor and alarm
High reliability
Typical Applications
Jacketed reactor vessels
Large rotary evaporators
Reaction calorimetry
Fermentation
Analytical instrumentation
Item | Description | |
ARB320-01 | ARB320-02 | |
Control Method | Cooling / heating automatic PID control PID | |
Heat Radiation | Peltier element(Liquid-Cooled) | |
Operating Temp | 0~60℃ (for Water ≥4℃) | |
Temp. stability | ±0.04︒C | |
Temp. distribution | ±0.04︒C | |
Cooling Capacity | 320W @T-bath=25℃;T-facility=25℃ | |
Heating Capacity | 700W @T-bath=25℃;T-facility=25℃ | |
Bath Liquid | Water : 4-60℃ Ethylene glycol-Water ,(≤50%) : 0-60℃ | |
|
Facility Liquid | Temp | 10~35℃ (No Dew Condensation) | |
Flow rate | 2~7 L / min | |
Pressure | ≤145 PSI (1 MPA) | |
port size | RC 3/8 | |
Size | Bath | W216xD216xH250mm Usable Dept: 220mm | W300xD350xH400mm Usable Dept: 370mm | |
|
Overall | W266xD376xH400mm | W350xD510xH550mm | |
Drain Port size | CPC Coupling PLCD 16006 | |
Power supply | 94~265 VAC, Single phase, 50/60HZ, 600W | |
Over Current Protection | Rated current, 10A | |
Serial COMM | RS485( RS 232 Option) | |
Alarm Output | High/Low temp alarm, Output cutoff alarm | |
Temp. Sensor | PT100, 3 wires | |
Ambient | Temp | 10~35℃ | |
Humidity | ≤85% RH | |
Weight | Approx. 20KG | Approx. 30KG | |