Model | TL-LC64 | TL-LC100 | TL-LC150 | TL-LC200 |
Inner chamber material | 400×400×400 | 500×450×450 | 600×500×500 | 650×460×670 |
External chamber material | 1440×1910×1825 | 1520×1990×2050 | 1620×2040×2100 | 1620×1980×2140 |
Temperature shock range | A:-40~+150℃, B:-55~+150℃, C:-65~+150℃ |
Cold storage temperature range in low temperature zone | 10℃~-50℃; -10℃~-65℃; -10℃~-75℃ |
Heat storage temperature range in high temperature zone | +50℃~+180℃ |
Impact reset time | About 5 mins |
The constant temperature time of high-low temperature shock | Above 30 min |
High temperature setting | +60℃ ~ +150℃ |
Low temperature setting | -65℃ ~ -10℃ |
Temperature fluctuations in the sample area | ±0.5℃ (at constant temperature) |
Temperature accuracy range | Setting accuracy: ±0.1℃, indication accuracy: ±0.1℃, resolution:
±0.1℃ |
Temperature variation | ±2°C (no load, at constant state) |
Temperature sensor | High-precision A grade PT100 temperature |
The refrigerant | R404a,R23 |
Chamber structure | With independent product test area, high temperature storage area,
low temperature storage area |
Inner chamber material | SUS304 stainless steel plate |
External chamber material | Cold rolled board baking varnish |
Insulation material | Rigid polyurethane foam + insulation cotton |
Circulation method | Multi-wing fan |
Heater | Nickel alloy electric heater |
Condenser | Shell-tube condenser(water-cooled) |
Evaporator | High-performance fin type with automatic load capacity adjustment,
which can be used for a long time under low temperature and high
humidity conditions without frosting |
Controller | 7-inch color LCD touch control screen + PLC module |
Program capacity | Usable program capacity: 120 groups at most; time setting: 99 hours
and 59 minutes per segment; |
Operating power | About 9~30KW (Automatically turn on heating and cooling system
according to PID calculation) |
Power supply | 380V±10%V; 50Hz; three-phase four-wire + grounding wire, protective
grounding resistance is less than 4Ω |
Safety equipment | Over-temperature protection, motor overload protection, compressor
high-voltage protection switch, compressor over-heat protection
switch, compressor over-current protection, over-voltage
under-reverse phase protection switch, leakage circuit breaker,
cooling water abnormal protection. |