China Electrical Appliance Testing Equipment manufacturer
Sinuo Testing Equipment Co. , Limited
Sinuo Testing Equipment Co., Limited
7
Home > Products > Environmental Test Chamber >

ISO 4892-1-2016 UV Weathering Accelerated Ageing Test Chamber

ISO 4892-1-2016 UV Weathering Accelerated Ageing Test Chamber

Brand Name Sinuo
Model Number SN882
Certification Calibration Certificate (Cost Additional)
Place of Origin China
Minimum Order Quantity 1
Price Customized
Payment Terms L/C, , T/T, Western Union, MoneyGram
Supply Ability 20 sets per month
Delivery Time 30 Days
Packaging Details Plywood
Applied Standards ISO 4892-1,ASTMD 4329
Light Source Q-lab UV-A 340
Wave Length Range 295-400nm
Chamber Temperature Range RT+5℃~85℃
Blackboard Temperature Range 40℃~80℃
Control Method PLC & touch screen
Detailed Product Description

 

ISO 4892-1-2016 UV Weathering Accelerated Ageing Test Chamber

 

 

Product Description

 

This equipment uses a fluorescent ultraviolet UV lamp to simulate the effect of sunlight, uses condensed moisture to simulate rain and dew, and puts the test material in a cycle program of alternating light and moisture at a certain temperature for testing.
It can detect paint, ink, paint, resin, plastic, rubber, printing and packaging, aluminum profile, adhesive, automobile, motorcycle industry, cosmetics, metal, electronics, electroplating, medicine, etc.
Used to simulate the harm caused by sunlight, rain, and dew.

UV can reproduce the hazards that occur outdoors for months to years in a few days or weeks. The hazard types include: fading, discoloration, loss of gloss, chalking, cracking, turbidity, bubbles, embrittlement, strength decline and oxidation.

 

The UV weathering accelerated ageing test chamber conforms to standards:

GB/T 14522-93 “Accelerated Weathering Test Method for Plastics, Coatings, Rubber Materials Used for Machinery Industrial Products”

ASTM G154 “Standard Practice for Operating Fluorescent Ultraviolet (UV) Lamp Apparatus for Exposure of Nonmetallic Materials1”

ISO 4892-1-2016 “Plastics-Methods of exposure to laboratory light sources-Part 1: General guidance”

ISO 11507 -2007 “Paints and varnishes - Exposure of coatings to artificial weathering - Exposure to fluorescent UV lamps and water”

ASTMD 4329-2005 “Standard Practice for Fluorescent UV Exposure of Plastic”

 

Technical Parameters

Inner Size D×W×H450×1170×500 mm

External Size

D×W×H

630×1350×1450 mm
Control SystemSiemens touch screen and PLC controller
Main componentsBrand ones: Schneider Omron electrical components
Light SourceLamp typeUVA-340
Wavelength range315nm (its peak value is 340nm)
Number of lampsTotal 12 pcs (8 are installed on the chamber, 4 for spares)
Lamp power40W/pc
Lamp brandQ-lab, American brand
Lamp lifeApproximately 1600 hours
Equipment FunctionTemperature, humidity, light, spray, condensation
Irradiance monitor: mounted irradiator, 0-5V signal output
Illumination adjustment: dimming system

Parameters

 

Temperature rangeRT+10℃~70℃
Light cycle45℃-80℃
Condensing cycle40℃-60℃
Humidity range≥95%R.H
Temperature fluctuation≤±0.5℃
Temperature uniformity≤±2℃
Temperature resolution±0.1℃
Irradiation intensity0.3-1.2W/㎡ adjustable
Black panel temperature40℃~65℃
Distance between lamps35mm
Distance between Sample and Lamp50mm
Spray time0~999 minutes can be preset on touch screen
Test duration0~99H59M, can be preset on touch screen
Standard test piece

Flat plate type,

W75 mm*L150 mm* T max. 1.5mm

(Prepared by client according to standard requirements)

Numbers of test piece

Approximately 40 pcs

Each side of the chamber can hang 20 pieces

Condensation (tap water or deionized water) or spray consumption8L/day
ControllerSiemens control system, can control and monitor the humidity, condensation, water spray, and irradiance on the touch screen. The irradiance can be tested continuously.
Power supply220V,50HZ, 2.0KW or match client's local power supply requirements
Structure and MaterialInner materialSUS304 stainless steel plate
Outer materialSUS304 stainless steel plate
ShapeTower shape, upper and lower two floors
Use EnvironmentTemperature15~35℃
Humidity50-70 %RH
Air pressure86~106Kpa

1. The ground is level and well ventilated
2.No strong vibration around the equipment

3. No direct sunlight or other heat source radiation

4. There is no strong air flow. When the surrounding air needs to flow forcibly, the air flow should not be directly blown onto the cabinet;
5.No strong electromagnetic field around the equipment
6.No flammable, explosive, corrosive substances and dust around the equipment
7. A space of 300mm should be reserved between the equipment and walls or utensils for operation, maintenance, and ventilation.

 

 

 

 

Product Tags: PLC Ageing Test Chamber   UV Weathering Accelerated Ageing Test Chamber   400nm Ageing Test Chamber  