Desktop LCR Digital Multimeter Bandwidth Frequency Adjustable VICTOR 4091C 10 Hz ~ 100 KHz LCR Meter Bench Top bandwidth
frequency Data retention adjustable 0.2% Accuracy 3.5 inch TFT
display Product description LCR digital bridge is designed using automatic balance bridge
principle component parameter analyzer, the tests of 10 hz ~ 100
KHZ bandwidth, frequency adjustable, 0.2% of the basic
classification function provided by the accuracy of measurement and
instrument for the vast majority of components and materials to
provide accurate and complete the survey and analysis, is widely
used in product research and development, components such as
incoming inspection, product online applications. Technical descriptionOutput signal: low effective, open collector output, Opto-couplers
isolation. Input signal: Opto-couplers isolation DC power supply voltage supported by power pin: 3V-25V. Product features - Ø Measuring frequency 10Hz- 100 KHZ, 1 Hz step continuously
adjustable
- Ø Test level 10 ~ 2000 mv, 1 mv step continuously adjustable
- Ø Support dc resistance (DCR), electrolytic capacitor measurement
- ØThe internal bias voltage output 10 mv - 1500 mv
- Ø 3.5 inch TFT display, five shows
- Ø Communications: USB Device, RS232 (or 485), Handle, GPIB
(optional), USB Host (optional)
- Ø Data record function (maximum, minimum, average)
- Ø Support SCPI communication protocol
- Ø Provide the system Settings, can according to need to configure
the Chinese or English, buzzer, screen brightness, etc
- ØBasic measurement accuracy of 0.2%
- Ø Manual and automatic range
- Ø With open circuit and short circuit calibration function
- Ø Comparator separation (5), alarm function
DCR measurement accuracy | Range | Display range | Accuracy Re | | 100MΩ | 20.00MΩ-99.99MΩ | 10.0%+20 count | | 20MΩ | 10.00MΩ-20.00MΩ | 5.0%+10 count | | 10MΩ | 4.000MΩ-9.999MΩ | 2.0%+5 count | | 4MΩ | 400.0kΩ-3.9999MΩ | 1.2%+3 count | | 400kΩ | 40.00kΩ-399.99kΩ | 0.3%+3 count | | 40kΩ | 4.000kΩ-39.999kΩ | 0.2%+2 count | | 4kΩ | 400.0Ω-3.9999kΩ | 0.1%+2 count | | 400Ω | 40.00Ω-399.99Ω | 0.2%+2 count | | 40Ω | 4.000Ω-39.999Ω | 0.3%+3 count | | 4Ω | 0.400Ω-3.999Ω | 1.0%+5 count | | 0.4Ω | 0.000Ω-0.399Ω | 3.0%+10 count |
Main technical indicators| Model |
4091C | | Testing frequency | 10~100KHz Continuously adjustable, a step of 1 Hz | | Display screen |
3.5” TFT LCD screen | | Number of display digits | Principal parameter: 5 digit Secondary parameter: 5 digit | | Measured parameter: | Principal parameter: L/C/R/Z Secondary parameter: X/D/Q/θ/ESR | | Measurement range | L:0.001µH~99999H, C:0.001pF~99.999mF, R:0.0001Ω~99.999MΩ | | Basic accuracy | 0.2% | | Measuring display speed | 2 time/s (slow), 4 times/s (medium), 8 time/s (fast) | | Internal bias | 0-1500mV adjustable, at a step of 1mV. | | Testing level | Six fixed level(0.1V,0.3V,0.6V,1V,1.5V,2V) | 0.1~2V adjustable, at a step of 1mV. | | Signal source output impedance | 30Ω,100Ω | | Calibration function | Open circuit calibration, short circuit calibration | | Screening function | The limit range of screening can be set to -50%~+50%, and the fixed
points are 1%, 5%, 10% and 20%. | | Comparator | 5 groups sorting,3 groups of qualified setting, one group of
unqualified setting, one group of auxiliary setting | | Interfaces | standard: RS232(or 485), USB Device, Handler ;optional :GPIB,USB
Host | | Others | Support dc resistance (DCR), electrolytic capacitor measurement
model,Adjustment of backlight brightness, Chinese and English are
optional |
General technical specifications - Power voltage: 220V.AC/50Hz ±10%,OR 110V.AC(optional)
- Power consumption: <10W
- Display: 3.5” TFT LCD, resolution 480*320
- Interfaces: standard: RS232(or 485), USB Device, Handler ;optional :GPIB,USB Host
- Service environment: 0°C-40°C
- Sizes: 265mm*105mm*305mm (L*W*H)
- Weight: 2.1 kg
|