Selling leadsDoppler Ultrasound Machine Repair Solutions Mindray DC-70 Dark Shadows In The Image Of The Echo Zone
... exists when switching the probe and replacing the interface. Prioritize the TR board, and it still exists after replacement. After replacing the ...
2025-06-18 18:36:55
|
...Ultrasound Machine Repair Cell Board 1. Repair time: 2-4 days 2. Warranty: 60 days 3. Fault phenomenon: All probes have no echo images, and the ...
2025-07-30 00:11:34
|
...Ultrasound Machine Repair DICOM Error Reported During Use 1. Repair time: 2-4 days 2. Warranty: 60 days 3. Fault phenomenon: Philips HD7 reports ...
2025-06-18 18:36:55
|
...Ultrasound Machine Repair Touch Screen Flower Screen 1. Repair time: 2-4 days 2. Warranty: 60 days 3. Fault phenomenon: The device can enter the ...
2025-06-18 18:36:55
|
...Ultrasound Machine Repair Service Unable To Enter The System After Booting Model: Siemens X300 Fault Phenomenon: cannot enter the system after ...
2025-07-30 00:11:34
|
...test and check the log, it is judged that the RXBF board is faulty. Maintenance result: After replacing the RXBF board, the equipment is normal. ...
2025-06-18 18:36:55
|
...the fault of the beamforming module, and it is judged that the RFM is faulty. 4. Solutions: After replacing the RFM, reboot the machine and the ...
2025-06-18 18:36:55
|
.... After replacing the ACB, it can be turned on and used, but the error 0022 is reported when the probe is switched, and the self-test AIM fails ...
2025-07-30 00:11:34
|
...path and blood flow, and all probes were faulty. Determine the RDBF fault. Maintenance result: After replacing the RDBF board, the equipment is ...
2025-07-30 00:11:34
|
...Ultrasound Machine Repair Report Error 201 When Booting Medical Equipments 1. Repair time: 2-4 days 2. Warranty: 60 days 3. Fault phenomenon: Enter ...
2025-06-18 18:36:55
|