Mindray IPM-9800 SPO2 Plate 050-00046-00 110mm×90mm×1.6mm
|
Detailed Product Description
|
Mindray IPM-9800 Monitor SPO2 Plate 050-00046-00 051-000058-00 Mindray IPM-9800 Monitor SPO2 Plate includes two models: 050-00046-00 and 051-000058-00. The 050-00046-00 focuses on signal processing, while the 051-000058-00 focuses on interface drivers. Together, they form a complete blood oxygen monitoring unit, the core hardware for accurate SpO₂ and PR measurements. Their materials and circuit design fully meet the stringent requirements of medical equipment for stability, interference immunity, and safety. Product Specifications
Core Functions
Frequently Asked Questions Q1: The device displays "SpO₂ signal weak." I've checked the probe
and it's fine. Is the oximeter board faulty? A1: Turn off the power and check if the power cord is securely
plugged in. After powering on the device, enter maintenance mode
and check the oximeter board self-test results. If it displays
"Signal amplification circuit abnormality" (fault code E04), the
oximeter board is faulty and should be replaced. Q2: Can the two oximeter boards be replaced separately? Or must
they be replaced together? A2: They can be replaced separately, depending on the fault type: ①
If "Probe cannot be recognized" or "Drive current abnormality"
occurs, it's likely that the 051-000058-00 board is faulty and can
be replaced alone. ② If "Oximeter calculation error is large" or
"Weak signal cannot be calibrated" occurs, it's likely that the
signal processing chip or filter circuit on the 050-00046-00 board
is faulty and should be replaced alone. ③ After replacement, power
on the device and perform "SpO₂ calibration" to ensure measurement
accuracy meets requirements. Q3: The blood oximeter board gets noticeably hot while operating
(approximately 65°C to the touch). Is this normal? A3: It's not normal. The normal operating temperature should be
≤60°C. Possible causes of abnormal heating: ① The heat sink may
have fallen off or the thermal grease may have dried out.
Disassemble the device to inspect the heat sink on the 050-00046-00
board and reapply thermal grease (0.5mm thick). ② The driver chip
is overcurrenting. Measure the probe LED current to see if it
exceeds 80mA. Product Images Ordering Information
|
||||||||||||||||||||||||
| Product Tags: Mindray IPM-9800 SPO2 plate patient monitor SPO2 sensor Mindray IPM-9800 replacement part |
Related Products
|
Mindray Patient Monitor Battery LI13I001A Suit For Ipm5 Ipm8 Ipm10 Ipm12 IMEC5 IMEC8 IMEC10 IMEC12 New |
|
051-000943-00 Patient Monitor Spo2 Board Mindray T1 IPM12 IPM10 IPM8 Imec8 Imec10 Imec12 |
|
Patient Monitor Printer Recorder TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series |
|
PHiliph MX400 MX450 MX Series Patient Monitor AC/DC IV2-FLEX ASSY-PWR AC / DC Power Supply 453564281221 |
|
P/N 453564271711 PHiliph MX400 Mx450 MX Series Patient Monitor Mainboard |
|
MX450 Patient Monitor Battery Board Power Supply Hospital Equipment |
Email to this supplier