China Used Patient Monitor manufacturer
China World Technology Medical Equipment Service Group
China World Technology Medical Equipment Service Group Win-win together!
6
Home > Products > Patient Monitor Parts >

PHiliph MP60 MP70 Patient Monitor Module Rack Bedside M8048A FMS-8

Browse Categories

China World Technology Medical Equipment Service Group

City: guangzhou

Province/State:guangdong

Country/Region:china

Tel:86-198-5481-5217

Contact Person:
Miss.Kiara
View Contact Details

PHiliph MP60 MP70 Patient Monitor Module Rack Bedside M8048A FMS-8

Brand Name PHiliph
Model Number M8048A FMS-8
Minimum Order Quantity 1pcs
Payment Terms Full before shipment
Delivery Time 3-5 working days
Packaging Details Carton
Product Name M8048A FMS-8 PHiliph MP60 MP70 Patient Bedside Monitor Module Rack
Brand PHiliph
Part no. M8048A FMS-8
Weight <3500g ,<7.7 lb
Weight*Height *Width <315mm*130mm*120mm (12.4*5.1*4.7inch )
Warranty 90 days
Condition New & Original
MOQ One piece
Payment Method T/T ,Paypal ,Western Union
Color White
Detailed Product Description

M8048A FMS-8 PHiliphhilip MP60 MP70 Patient Bedside Monitor Module Rack

 

 

 

 

PHiliphhilip FMS module rack

 

MAX weight:<3500g ,<7.7 lb

Weight*Height *Width:<315mm*130mm*120mm (12.4*5.1*4.7inch )

 

Flexible Module Server Troubleshooting

 

SymptomsCause of FailureFailure Isolation and Remedy
Prompt Message “Unrecognized Measurement Module in slot s” is issuedAn unsupported module has been plugged into the Flexible Module ServerUnplug the unsupported module
Prompt message “Measurement Module in slot n is currently ignored” is issuedToo many modules of the same kind have been plugged into the Flexible Module ServerUnplug module in slot n
Red Error LED stays onUnrecoverable hardware selftest error:Try connecting the MMS directly to the MSL cable. If the measured value is displayed on the screen, the fault is in the FMS
MSL cable defectiveIf no measurements are displayed when the MMS is directly connected to the MSL cable, replace the MSL cable
Flex connector from main board to MSL defectiveReplace connector
CPU module defectiveReplace CPU module
Red Error LED flashesHardware selftest errorIf system comes up, check status log. Otherwise see above
Flexible Module Server LEDs OK, Front Measurement Module not recognized (silent or INOP)Damaged measurement module or measurement module connectorReplace Measurement Module
 No Front-End power because MSL voltage from the monitor is too high or too low

1. Try a new MSL cable. Replace if failure is rectified.

2. Replace mother board

 Mother board or connector on Mother Board defectiveIf the voltage is within range, or if the connector is visibly damaged, replace the motherboard

 

 

 

 

PHiliphhilip FMS/MMS Module Lock

 

 

The following parts and hardware are included in this mounting kit

 

Item #DescriptionQty
1Main Enclosure1
2Support Bracket1
3#8-32 x 1/4'' Button Head Machine Screw, Tamper-proof4
45/32'' Hex Wrench, Tamper-proof1

 

 

Mounting Notes: The bracket is designed to accommodate FMS (handle on both ends) or MMS extension (handle on one end). Installation is only shown on FMS with MMS extension.

 

 

Installing the Module Lock

 

1. After installing the required modules, slide the FMS into the main housing as shown below. Installation Note: It is not necessary to remove the previously installed mounting clips from the FMS to install the module lock.


 

 

2.Slide the support bracket over the MMS extension and into the main housing as far as it will go. Note the slot on the bracket for mounting the module lock on the second handle.


 

3.Using the 5/32'' tamper-resistant hex key provided, install four (4) #8-32 x 1/4'' tamper-resistant screws as shown below.

Installation Notes: Module locks are designed to provide space for individual modules to be disconnected without removing them from the server.

 

 


 

 

 

 

Product Tags: MP70 Patient Monitor Module Rack   M8048A Patient Monitor Module Rack  
Related Products
Email to this supplier
 
From:
Enter your Email please.
To: China World Technology Medical Equipment Service Group
Subject:
Message:
Characters Remaining: (80/3000)