10 Gbps Full Rate CDR Adaptive Rate 2/4 Frequency
Division
U9710 CDR (Clock Recovery Module ) is mainly used to provide
clock recovery signal in the test environment without triggering
clock, so as to use the sampling oscilloscope to test the optical
eye diagram of the device. The typical application is the testing
of optical network equipment or the BOB prodUt in
PON application. Because the photometric device itself has no
frequency division clock output, it is necessary to recover the
clock signal to trigger the sampling oscilloscope when testing the
optical eye diagram.
Full-rate clock recovery
Support clock signal recovery at rates of 10Gbps and below (such as
1.25Gbps, 2.5Gbps, 6.25Gbps, etc.), covering common communication
standards (such as 10GBASE-R, 10G Fibre Channel, etc.). Some
high-end models can be expanded to 25Gbps or higher.
Low jitter and high precision
The jitter of the restored clock is usually less than 1ps
(picosecond level), ensuring the accuracy of the test results.
NO. | technical index | maximum value | minimum value | unit |
1 | Data rate range | 11.3 | 0.155 | Gbps |
2 | Internal signal loss detection (LOS) | 128 (overload) | 5(sensitivity) | mV |
3 | Power supply voltage (external power supply input voltage) | 5.5 | 4 | V |
4 | Operating temperature | 85 | -40 | °C |
5 | Clock rise time (OC-192) (20%~80%) | 33.1 | 22.2 | ps |
6 | Clock down time (OC-192) (80%~20%) | 33.7 | 23.9 | ps |
7 | Clock frequency (10.3125Gbps rate, 2 frequency division) | 2.578 | 2.578 | GHz |
8 | Output jitter (10.3125Gbps) | 1 (typical value) | ps |
The 10Gbps full-rate clock recovery instrument is a key device used
for high-speed communication testing. Its core function is to
extract clock signals from high-speed data streams, providing a
stable trigger clock for test equipment (such as sampling
oscilloscopes), thereby achieving precise analysis of signal
quality.