The Wavelength Range Of O Band Laser Scanning Spectrometer Is 1260~1360nm

Brand Name YOUCii
Model Number U8512O
Place of Origin CHINA
Minimum Order Quantity 1pcs
Payment Terms L/C,D/A,D/P,T/T,Western Union,MoneyGram
Certification CU,IOS
Supply Ability 20 pcs+per month
Delivery Time 20 work days
Packaging Details Carton
Price negotiate a price
Wavelength range 1525.00 ~ 1568.00 nm
Wavelength resolution 1.0 pm
Wavelength repeatability ± 3 pm, typ. ±1 pm
Power stability ±0.005dB (15Min) ±0.05 dB ( 24 hours)
Power linearity ± 0.3 dB
Relative noise intensity Typical value < -135 dB
Detailed Product Description

The wavelength range of O-band laser scanning spectrometer is 1260~1360nm


   The U8512 laser scanning spectrum analyzer is a passive device testing and spectrum analysis system controlled by PC. It is mainly used for coarse WDM components, high-density WDM components, Interleaver, AWG, WSS, OPM, DPSK components and other general optical fiber optical measurement and application. The tunable laser light source is integrated in the system, and with the optical power meter, it can be a tunable laser light source test system platform with high wavelength accuracy and fast scanning function.


   Equipped with PC test software, U8512 can flexibly build test templates and create test reports according to customer needs. It is a good test platform in the process of adjusting and monitoring the optical path of passive devices. The fast scanning function of U8512 can directly display the spectrum of the product, and the built-in tunable light source and external power meter can be used separately or jointly to realize the function of scanning spectrum analyzer.



Model number

U8512O

Wavelength range

1525.00 ~ 1568.00 nm

Output power

13 dBm

Wavelength resolution

1.0 pm

Absolute wavelength accuracy

±10 pm, Typ. ±5 pm

Relative wavelength accuracy

± 5 pm, Typ. ±2 pm

Wavelength repeatability

± 3 pm, typ. ±1 pm

Wavelength stability

±5 pm (typ., 24 hrs at Constant Temperature)

Tuning speed

≤ 2 ms per step

Power stability

±0.005dB (15Min) ±0.05 dB ( 24 hours)

Power repeatability

± 0.05 dB

Power linearity

± 0.3 dB

Power flatness

0.3 dB; Typical value 0.05dB.

Edge mode rejection rate

≥ 40 dBm

Relative noise intensity

Typical value < -135 dB

Power adapter

AC 100 ~ 240 V , 48 - 66 Hz,100 VA max.

Display interface

PC control interface

Overall dimension

245 mm W, 105 mm H, 320 mm D