1. Product Brief Introduction:
The HQ-LWQ gas turbine flow meter adopts advanced microprocessor
technology, possessing advantages such as powerful functions, high
calculation accuracy, and reliable performance. Its main technical
indicators reach the advanced level of similar foreign products. It
is an ideal instrument for gas metering in petroleum, chemical,
power, metallurgical, industrial and civil boilers, as well as for
metering urban natural gas, gas pressure regulating stations, and
gas trading. Its working principle is as follows: When gas enters
the flow meter, it is first accelerated by a special rectifier.
Under the action of the fluid, the turbine overcomes resistance
torque and friction torque and begins to rotate. When the torque
reaches equilibrium, the speed stabilizes. The turbine speed is
proportional to the gas flow rate. The rotating signal disk
periodically changes the sensor's magnetic resistance, thereby
causing the sensor to output a pulse signal proportional to the
flow rate.
2. Product Features:
(1) It adopts imported German precision bearings, ensuring high
accuracy, good stability, and a wide rangeability (20:1).
Small-diameter flow meters require no lubrication for five years
under normal operating conditions, while large-diameter flow meters
only require occasional lubrication, making them easy to use.
(2) The carefully designed flow channel structure avoids airflow
between the bearings, improving the medium adaptability of the
turbine flow meter.
(3) The unique reverse thrust structure and sealing structure
design ensure long-term reliable bearing operation.
(4) It uses a magnetoresistive element instead of a magnetic
sensing coil, avoiding magnetic attraction, improving detection
sensitivity, further reducing the starting flow rate, and enhancing
product stability and reliability.
(5) The independent mechanism design ensures good
interchangeability and convenient maintenance.
(6) It integrates temperature, pressure, and flow sensors with an
intelligent flow totalizer, automatically tracking and correcting
the temperature, pressure, and compressibility factor of the
measured gas, directly measuring the standard volumetric flow rate
and total volume of the gas.
(7) The main performance indicators reach international advanced
levels and comply with ISO9951 standards. 8. Employing advanced
low-power technology, it can operate with either internal or
external power supplies; the internal battery can last for over
five years.
(8) The HQ-LWQ model boasts powerful functions, offering four
compensation methods, three pulse signal outputs, three historical
data recording methods, and two standard current signal outputs.
(9) It can form a network communication system via an RS485
interface, facilitating automated management. The RS485
communication protocol conforms to the MODBUS standard.
(10) The meter head can rotate 180° freely for easy installation.
3. Technical Specifications:
Main Technical Parameters of the HQ-LWQ Gas Turbine Flowmeter
3.1 Model Specifications and Basic Parameters
| DN(mm) | Model |
| Range(m³/h) | Expend | pressure rating(Mpa) | High pressure(Mpa) | Installation |
| (m³/h) |
| DN25 | HQ-LWQ-25 | S | 2.5-25 | W | 4-40 | 1.6 | 2.5,4.0 | Flange(thread) |
| DN40 | HQ-LWQ-40 | S | 5-50 | W | 6-60 | 1.6 | 2.5,4.0 | Flange(thread) |
| DN50 | HQ-LWQ-50 | S1 | 6-65 | W1 | 5-70 | 1.6 | 2.5,4.0 | Flange |
| S2 | 10-100 | W2 | 8-100 |
| DN65 | HQ-LWQ-65 | S | 15-200 | W | 10-200 | 1.6 | 2.5,4.0 | Flange |
| DN80 | HQ-LWQ-80 | S1 | 13-250 | W | 10-160 | 1.6 | 2.5,4.0 | Flange |
| S2 | 20-400 |
| DN100 | HQ-LWQ-100 | S1 | 20-400 | W | 13-250 | 1.6 | 2.5 | Flange |
| S2 | 32-650 |
| DN125 | HQ-LWQ-125 | S | 25-700 | W | 20-800 | 1.6 | 2.5 | Flange |
| DN150 | HQ-LWQ-150 | S1 | 32-650 | W | 80-1600 | 1.6 | 2.5 | Flange |
| S2 | 50-1000 |
| DN200 | HQ-LWQ-200 | S1 | 80-1600 | W | 50-1000 | 1.6 | 一 | Flange |
| S2 | 130-2500 |
| DN250 | HQ-LWQ-250 | S1 | 130-2500 | W | 80-1600 | 1.6 | — | Flange |
| S2 | 200-4000 |
| DN300 | HQ-LWQ-300 | S | 200-4000 | W1 | 130-2500 | 1.6 | 一 | Flange |
| W2 | 320-6500 |
3.2 Measured Media
Natural gas, city gas, and various other fuel gases, alkanes, and
industrial inert gases.
3.3 Operating Conditions
◆ Ambient temperature: -30℃~+60℃; ◆ Medium temperature: -20℃~+80℃;
◆ Atmospheric pressure: 70kPa~106kPa; ◆ Relative humidity: 5%~95%.
3.4 Electrical Performance Indicators
3.4.1 Power Supply and Power Consumption
a. External
power supply: +24VDC ±15%, suitable for 4mA~20mA output, pulse
output, RS485, etc.;
b. Internal
power supply: 1 set of 3.6V lithium batteries, can be used
continuously for more than five years.
3.4.2 Pulse Output Mode (LWQ type can be
selected from one of the following three options)
a. Working
condition pulse signal, opto-isolated amplified output, high level
amplitude ≥20V, low level amplitude ≤1V.
b. Frequency
signal proportional to the standard volume flow rate, opto-isolated
amplified output, high level amplitude ≥20V, low level amplitude
≤1V.
c.
Calibration pulse signal (compatible with IC card valve
controller), high level amplitude ≥2.8V, low level amplitude ≤0.2V,
the volume represented by a unit pulse can be set in the range of
0.01m³~10.00m³.
3.4.3 RS485 Communication (using opto-isolated
RS485 communication module), can be directly networked with a host
computer or secondary instrument to remotely display current data
and historical records.
3.4.4 4mA~20mA standard current signal (using
opto-isolated standard current module); two-wire or three-wire
system.
3.4.5 Control Signal Output
a. Upper and
lower limit alarm signals (UP, LP): Opto-isolated open collector
(OC) output, in normal state the OC gate is off, in alarm state the
OC gate is on, maximum load current 50mA, working voltage
+12VC~+24VDC.
b. Valve
closing alarm (BC) and battery undervoltage alarm (BL) output (for
IC card controller); logic gate circuit output, normal output is
low level, amplitude ≤0.2V; alarm output is high level, amplitude
≥2.8V, load resistance ≥100kΩ.
3.5 Real-time data storage function
a. Start/stop record: Records of the most
recent 1200 start/stop times and total volume.
b. Daily record: Records of the date,
temperature, pressure, standard volume flow rate, and total volume
at midnight for the most recent 920 days.
c. Timed interval record: 920 records of 8
periods of time, temperature, pressure, standard volume flow rate,
and total volume at timed intervals.
3.6 Explosion-proof rating: Flameproof type Exd II BT4.
Intrinsically safe type Exia II CT4.
3.7 Protection class: IP65
4. Dimension:
Dimensions of the HQ-LWQ Gas Turbine Flowmeter
5. Selection:
HQ-LWQ Gas Turbine Flowmeter Selection
4.1 Applicable scope
a. Applications requiring a flow range ratio
less than 20:1 (see Table 1), and with high requirements for
starting flow rate.
b. No frequent flow fluctuations with short
intervals and large amplitude.
c. Can measure natural gas, city gas,
compressed air, nitrogen, etc.
4.2 Specification determination
Calculate the flow range under working conditions based on the gas
supply flow range, medium pressure, and temperature under standard
conditions (refer to the selection of vortex flowmeters).
4.3 Pressure loss of the flowmeter
Calculate the maximum pressure loss △Pmax of the flowmeter at the
maximum flow rate under working conditions using the following
formula (1). The maximum pressure loss of the flowmeter must
satisfy condition (2) to ensure normal operation. If the pressure
loss does not satisfy formula (2), a larger size should be
selected.
a. Pressure loss can be calculated using
the following formula:
P: Gas density under standard conditions (20℃, 101.325 kPa);
△Pomax: Pressure loss at maximum flow rate when the medium is
normal dry air (density 1.205 kg/m³) (obtained from Table 1);
Pa: Local atmospheric pressure (kPa); Q: Operating flow rate
(m³/h);
Qmax: Maximum operating flow rate of the instrument (m³/h); Pg:
Gauge pressure of the medium (kPa).
Pn: Standard atmospheric pressure (101.325 kPa); Tn: Standard
temperature (293.15 K);
Tg: Temperature under operating conditions of the medium (273.15 +
t); where t is the operating temperature of the medium (°C);
Zn, Zg: Gas compressibility factors under standard and operating
conditions, respectively.
b. The pressure loss should satisfy the
condition:
P1 - △Pmax ≥ Lmin................... (2)
Where: P1: Minimum operating pressure of the medium at maximum flow
rate;
△Pmax: Maximum pressure loss of the flow meter at maximum flow rate
under operating conditions;
PLmin: Minimum inlet pressure required for the gas appliance.
Selection Table
| HQLWQ | Gas Turbine Flow Meter |
| Range | □ | DN25-DN300 |
| Type | A | Gas Turbine Flow Meter |
| B | Wireless remote transmission type gas turbine flowmeter |
| C | Solar-powered gas turbine flowmeter |
| Measuring medium | A | Conventional gas |
| B | Combustible gas |
| C | Etchant gas |
| Medium temperature | A | 0-60℃ |
| B | 60-100℃ |
| Compensation methods | A | Temperature compensation |
| B | Pressure compensation |
| C | Temperature and pressure compensation |
| D | NO |
| Medium pressure | A | 0-1MPA |
| B | 1-1.6MPA |
| C | 1.6-2.5MPA |
| D | 2.5-4MPA |
| Output signal |
| NO |
| A | HART |
| B | RS485 |
| C | 4-20MA |
| Valve body material | A | NO |
| B | 304SS |
| C | 316SS |
| Installation | A | Flange |
| B | Thread |
| C | Clamp |
| Accuracy | A | 1.00% |
| B | 1.50% |
| Explosion-proof method | A | No explosion-proof |
| B | Explosion-proof (EX) |
| Power supply | A | 24VDC |
| B | Battery |
6. Installation: