Upgrade Endoscopy System Socket Unit For Olympus CLV-180 Light Source
|
|
Socket unit for Olympus CLV-180
► Brief Description
The Socket Unit for the Olympus EVIS EXERA II CLV-180 Xenon Light Source is a high-precision, heavy-duty electrical and mechanical interface designed to house and power the 300W Xenon short-arc lamp (typically the MD-631 or MD-631 equivalent). Positioned inside the lamp compartment of the light source, this unit acts as the critical bridge between the device's high-voltage ignition circuitry and the xenon bulb. It is engineered to withstand extreme thermal stress, deliver the massive voltage spikes required for arc ignition, and maintain precise optical alignment of the bulb to ensure optimal light transmission through endoscopic fiber-optic bundles.
►Features
► Specification
► FAQ
What causes a CLV-180 lamp socket to burn out or develop scorch marks?
The most common cause is poor mechanical contact (loose screws). If the hex screws securing the lamp assembly into the socket are not tightened adequately, it creates electrical resistance. Because a 300W Xenon lamp draws high current, this resistance generates extreme localized heat, leading to electrical arcing, pitting, and eventual melting of the socket contacts.
How to know if need to replace just the Xenon bulb or the entire socket unit?
Inspect the interface during a lamp change:
► More service
Spare parts for Wolf camera head, such as CCD, cable, cover, sheath/strain relief, etc. Camera system brand by OLYMPUS, WOLF, STORZ, Stryker, Smith&Nephew, ScanMed, etc.
Main models of camera head:
► About Us
CareUtech is the global provider of medical equipment and medical equipment repair service, as well as offering spare parts. We focus on repair and repair training for monitor, Flexible & Rigid Endoscope, ultrasound, camera system, dynamical system and surgical instrument.
Shipping and packing
► What we can do
|
||||||||||||||||||||
| Product Tags: Light Source Socket unit Olympus CLV-180 Socket unit CLV-180 Socket unit |
|
Olympus MAJ-1817 Xenon Lamp For Endoscopic Cold Light Source |
|
Touch Screen for Stryker Core Console Powered Instrument Driver |
|
Power Cable For Stryker 1288HD Camera Head |
|
Upgrade Endoscopy System Socket Unit For Olympus CLV-180 Light Source |
|
KEYPAD FOR STORZ TC200 CAMERA CONTRONER for Storz TC200 Camera Control |
|
ELECTRIC SOCKET FOR FUJINON VP-4400HD ENDOSCOPY PROCESSOR |