Digital Integrated Low-range 100NTU Turbidity Sensor For Drinking
Water >>Overview The turbidity sensor is based on the infrared scattering light
technology, that is, the infrared light emitted by the light source
will scatter through the measured sample during the transmission
process, and the intensity of the scattered light is directly
proportional to the turbidity. The turbidity sensor sets the
scattered light receiver in the 90° direction, and obtains the
turbidity value by analyzing the intensity of this set of scattered
light. It can be widely used in turbidity monitoring in sewage
water plant, water plant, water station, surface water, industry
and other fields. >>Technical Parameter Power/Outlet | 9~36VDC/RS485 MODBUS RTU | Measurement mode | 90°IR scattered light method | Dimensions | 110*80*80mm | Waterproof rating | IP65 | Measurement range | 0.001-100.00NTU | Measurement accuracy | 0 – 40 NTU, ±2%;40 – 100 NTU,±5% | Pressure resistance | ≤0.2Mpa | Flow velocity | 300ml/min≤X≤500ml/min | Measuring temperature | 0-45℃ | Calibration | Standard liquid calibration, water sample calibration | Cable length | Standard 1m,can be extended | Thread | Flow-through |
>>Dimension And Installation Install the sensor in place to ensure representative measurements.
Install the sensor in an easily accessible position to facilitate
regular cleaning and maintenance of the sensor. Install the sensor
near a representative, good sampling site: - Inlet water pipe, outlet water pipe, cleaning drainage pipe users
self-allocation, caliber for PE 3 in charge;
- It is recommended to add pressure reducing valve and valve to
control the front end of water inlet pipe.
|