Volume, size and weight |
Nominal inside volume | ( 150L) |
Inner box size | W(500 )mm × H( 600)mm × D(500 )mm |
Dimensions | W( 1070)mm × H( 1460)mm × D( 1070)mm |
Weight | ( 150)KG |
Performance |
Test environmental conditions | Under a condition of without sample inside the test chamber, the
environment temperature is + 5 ~ + 28 ℃, relative humidity ≤85%, |
Test method | GB/T 5170.2-2008 Temperature test equipment GB/T 5170.5-2008 Humidity test equipment |
Temperature range | (-20 )℃~+150℃ |
Control accuracy | Temperature:±0.2℃ (It means the difference between the controller set value and
measured value of the controller) |
Temperature fluctuation | ≤0.5℃(Temperature fluctuation is half of the difference between the
center of the highest measured temperature and the minimum
temperature) |
Temperature error | ≤±1℃(The average temperature of the studio temperature controller’s
display value minus the average temperature of measured
center) |
Temperature uniformity | ≤2.0℃(Temperature uniformity is the arithmetic mean value of the
difference between the measured maximum temperature and the minimum
temperature in each test) |
Temperature rise time | 3℃/min(Nonlinear load) |
Cooling time | 0.75~1℃/min(Nonlinear load) |
Work noise | A sound level≤70dB(A) (At environment temperature of 25 ℃, measured in the anechoic
chamber of less echo; adopt A-weighting, the average value of test
8 points; each test point level from the source of the noise is 1
meter ,height from the ground is 1 meter) |
Structure |
Thermal insulation structure | 1. Outer material: high-quality sand surface stainless steel plate 2. The inner wall material: high-quality mirror surface stainless
steel plate 3. Cabinet insulation material: rigid polyurethane foam + glass
fiber |
Chamber standard configuration | 1. Observation window: transparent electric membrane hollow glass 1
pc (on the door) 2. Lead hole: φ50mm 1 pc (In the left side of the box) 3. Preparation of the door interior lighting (Longevity and high
efficient energy-saving lamps) 4. Mobile casters: 4 pcs 5. Fixed feet cup: 4 pcs (the height is adjustable) 6. Sample frame: two layers of stainless steel sample holder, 8
hooks, load-bearing (uniform): 25kg / layer (The sample inside does not exceed the cumulative total load:
100kg) 8. PC communication software CD, U disk interface, RS23
communication interface, misuse hardware lock, communication cable
1 pc |
Door | Single open hinged door (facing the cabinet, hinge in the left
side, the handle in the right side), with observation windows,
lighting, window / door anti-condensation electric device |
Control panel | The controller display screen, U disk interface, RS232 computer
communication interface, misuse hardware lock, power switches,
emergency stop switches, status indicator light and audible warning
device |
Machine room | Refrigeration unit, water pan, drain hole Condenser fan, condenser inlet |
Control cabinet for power distribution | IO board, machine transformers, ballasts, intermediate relays, time
relays, solid state relays, AC contactors, thermal relays, fuses,
owe anti-phase relays, air switch |
Heater | 1. Nichrome strip heater 2. Heater control way: non-contact and periodic
pulse-width-modulated, SSR (solid state relay) 3. Heater power: 2.5KW |
Cord hole and drainage hole | At the back of the box |
Refrigeration system |
Working mode | 0℃~-40℃ Model adopts single-stage compression,-40℃~-70℃ Adopt
binary cascade refrigeration way |
Refrigeration compressor | Fully enclosed low noise rotary compressor |
Evaporator | Outside fins internal thread tubular heat exchanger (Also used as
dehumidifier) |
Condenser | Air-cooled: Outside fins internal thread tubular heat exchanger |
Throttling device | Expansion valve + capillary |
Plate heat exchanger (Using cascade refrigeration system) | Stainless steel brazed plate heat exchanger |
Refrigerator control mode | 1. Control system automatically adjusts the best energy-saving
refrigerator running condition according to the test condition 2. Evaporator cooling capacity is switched by the control system
drive the solenoid valve 3. The compressor’s return air cooling circuit |
Refrigerant | R404a/R23(Ozone depletion indices are 0, R23 adopted by the cascade
refrigeration system) |
Electrical control system | |
Controller Model | SAMWON TEMI880 |
Displayer | TEMI880:5.7 inch,640×480 dot matrix,TFT color LCD displayer |
Operation mode | Program mode, setting mode |
Setting mode | TEMI880 Chinese & English menu (Freely selectable), touch
screen input |
Program capacity | TEMI880(touching-type) 1.Procedure: Maximum 100 groups 2. Number of segments: Maximum 10000 segments 3. Number of cycles: Maximum can be infinite loop |
Setting range | Temperature: According to the temperature operating range of the
device to adjust |
Resolution | Temperature: 0.1 ℃; Time: 1min; Humidity: 0.1% RH (temperature and
humidity test equipment) |
Input | PT100 Platinum thermal resistance |
Communication features | U disk interface, RS-232 interfaces, misuse hardware lock,
complimentary computer communications software |
Control mode | 1. Anti-integration saturation PID 2.BTC balanced temperature control mode + DCC intelligent cooling
control + DEC intelligent electrical control (temperature test
equipment) 3. BTHC balanced temperature and humidity control + DCC intelligent
cooling control + DEC intelligent electrical control (temperature
and humidity test equipment) |
Curve recording function | With battery protection RAM, setting and sample value and sampling
time can be saved, the maximum recording time is 180 days (when the
sampling period is 60S) |
Ancillary functions | 1. Hint functions of fault alarm and causes, treatment 2. Power-off protection function 3. Upper and lower limit temperature protection 4. Calendar timing function (automatic start and automatic stop
running) 5. Self-diagnosis function |
Software using environment (user-supplied) | IBM PC compatible machine,more than PⅡCPU,more than 128M
memory,with RS-232 communication interface |
Temperature and humidity measurement | Temperature: PT100 platinum thermal resistance Humidity: dry and wet bulb thermometer method (damp-heat type only) |
Environmental conditions | 1.Temperature:5℃~35℃ 2. Relative humidity:≤85% 3. Air pressure:86kPa~106kPa |
Maximum current Maximum power | (40)A (5.5)KW (The test data obtained at ambient temperature + 25 ℃, after one
hour no sample, stable operating conditions) |