Three Zone Thermal Shock Test Chamber / High Low Temperature Testing Equipment
|
|
Three Zone Thermal Shock Test Chamber / High Low Temperature Testing Equipment
1. Description: Environmental Test Chamber Degree in an instant by a high temperature and extremely low temperature environment can tolerate continuous, that is, in the shortest time test chemical changes or physical damage due to its thermal expansion and contraction caused. Suitable objects include metal, plastic, rubber, electronics ... and other materials can be used as a basis for improvement of their products or reference
2.Thermal Shock Chambers of the Structure:
. cable threading hole: in the upper part of the sample holder with the top of the housing is provided with a high-temperature test slide cable conduit.
. sample holder bearing: ≥30Kg.
3. Specifications :
4.Shock Chamber Works As Follows: . low temperature storage tanks: tank temperature is determined by the state of the duct heater, evaporator and the working status of the fan. After the expansion valve throttling the refrigerant flowing into the evaporator in the working chamber, the working chamber of heat absorption and gasification, to reduce the temperature of working chamber; gasified refrigerant is sucked and compressed to a high temperature, high pressure gas into the condenser compressor It is condensed into liquid, and then the screening program, and finally through the expansion valve throttling, re-entered the studio evaporator endothermic gasification and then compressed by the compressor suction
. high-temperature storage tanks: a central controller temperature sensing signal from the detection instant, and the set temperature signal is compared, the comparison signal obtained by the instrument PID logic circuit output signal to control the solid state relay is turned off or adjusting the heating time ratio output power level, so as to achieve the purpose of automatic temperature control.
. temperature shock test chamber: the meter automatically controlled high temperature valves, low or high temperatures between the storage chamber switched respectively to form a closed air circulation system and low temperature box or boxes, quickly reached the target temperature test
The Air Control:
5.Notes
. If the high-temp exposure range lower limit + 60°C is required or the low temp exposure range lower limit -60°C, please select the “ambient-temperature exposure” option
. If the temp fluctuation need to below ±1.8°C please select the “Low temp fluctuation package”
. Temperature heat-up/pull down time are applicable only during independent chamber operation
. If the automatically operated sliding door required, please select the “Automatically door package”
. Air-cooled condenser / Water-cooled condenser both available
. Apply on water-cooled condenser only
6. Safety Devices . Hot chamber overheat protection switch . Cold chamber overheat protection switch . Air circulator overload alarm . Refrigerator high/low pressure protector . Compressor temperature switch . Air pressure switch . Fuse . Water suspension relay ( water-cooled specification only) . Compressor circuit breaker . Heater circuit breaker . Test area overheat/overcool protector . Air purge valve
7.Control (Color LCD interactive touch-screen system)
Operation and settings simplified by the use of a touch-screen LCD display (instructions displayed on-screen). At-aglance confirmation of test patterns, test area temperatures, temperature cycles, and trend graph displays.
8.Maintenance Radiator Thermal Shock Test the refrigeration unit (condenser) should be regularly maintained and kept clean. Sticky dust the condenser and the compressor will generate false alarms tripped high pressure switch, the condenser should be regularly monthly maintenance, use a vacuum cleaner to use after a hard brush dust suction power or attachment of the condenser cooling mesh brushes or high pressure gas nozzle blown clean of dust.
Packaging & Shipping Simulation Animatronic Customer Triceratops are covered with air bubble film before put them into the wooden case, which not only has good shock absorption, impact resistance, heat sealing and also has the advantages of nontoxic, odorless, moisture corrosion, good transparency etc.
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Product Tags: environmental testing equipment environmental test chamber | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Constant High Low Temperature Humidity Chamber Professional For Electron And Electrical Equipment |
|
PID Intelligent Adjust Environment Test Chamber Constant Temperature Humidity HD-E702 Climate Chamber |
|
Baking Paint Type Constant Temp Humid Test Chamber Temperature Humidity Chamber Professional |
|
Temperature And Humidity Test Chamber With Environmental Test Systems |
|
100L Programmable Automatic Touch Screen Paint Type Temperature And Humidity Test Chamber |
|
Simulated Environmental Automatic Water Cooling Xenon Accelerated Aging Chamber |