Name Model number | Programmable humidity and temperature Chamber KJ-2091 |
| |
2.1 Scope of application | This equipment is applicable to the environmental simulation and
reliability tests of high and low temperature constant and
gradient, hot and heat test and other products, parts and
materials. |
| |
3.1 Effective volume | 150L | 225L | 408L | 800L |
3.2 Internal box dimensions | 50×60×50cm | 40×75×60cm | 60×85×80cm | 100×100×80cm |
3.3 Test box dimensions of outer box | 102×146×116cm | 102×162×126cm | 113×172×148cm | 156×186× 140cm |
|
| Referring to when the air-cooled ambient temperature is + 25℃ and
water-cooled water temperature is + 27℃, with no test load (partial
parameters according to notes) |
- Temperature / humidity range
| -40℃ to + 150℃ / 20% (limited temperature section, see area
diagram) to 98%RH |
- Temperature fluctuation degree
| ≤±0.3℃ |
- Temperature / humidity deviation
| ≤±2℃ / ≤±3%RH |
| ≤2℃ |
| Average non linear ≥ 3℃ / min empty load |
| Average non linear ≥ 1℃ / min empty load |
- Lift and temperature is overcharged
| ≤±2℃ |
- The inner box prevents condensation
| Special sheet metal structure and dew protection device are used on
the top of the box to prevent short circuit during electrical parts
testing |
| Noise ≤ 70dB (measured by the noise measuring device at 1m from the
equipment gate). Noise reduction scheme: 1. main noise sources such as compressor with vibration reduction
spring and damping pad can reduce noise, reduce equipment damage
factors due to vibration, and extend the service life of equipment.
The frozen pipe is fixed by a clamp to effectively reduce vibration The internal wall of 2. refrigeration unit combined with high
quality absorbing cotton can effectively reduce the noise effect of
refrigeration unit. Add vibration absorber hose outside of connecting the 3. compressor
and the pipe. |
4.11. Meet the test standards | GB/T2423.1-2008 test A: low temperature (part) GB/T2423.2-2008 test B: high temperature (part) * The GB/T2423.3-2008 test of Cab: is constant, hot and humid The GB/T2423.4-2006 test of Db: became hot and humid GB/T2423.34-2005 test Z/AD: temperature and wet combination GB /T2424.2—2005 Hot and heat test guidelines GB /T2423.22—2002 test N: temperature change. GJB 150.3-2009 high-temperature test GJB 150.4-2009 low-temperature test GJB 150.9-2009 hot & heat test GB/T 5170.2-2017 GB/T 5170.5-2008 GB/T 2423.50-2012 |
| Note |
| Matte stainless steel plate, 1.5mm thick | |
| Thickness of SUS#304 stainless steel plate thickness 1.2mm,
corrosion resistance, all joints TIG are welded continuously to
ensure no leakage after long-term test. The bottom plate is
strengthened with the overall explosion-proof function | |
| Thickness of 150mm hard polyurethane foam and glass cotton, flame
and fire resistance, meet the environmental protection requirements | |
| Left single door is open in type | |
| Single door is attached with heat and anti-sweat vacuum large glass
observation window, during any test, the surface without frost and
condensation phenomenon observation window. | |
| With 2 layer height adjustable test shelf of joist type bearing
over 30KG | |
| Original imported silicone sealing strip | |
| A 100mm diameter round hole with cover and soft plug is left on the
left side of the equipment. | |
| Common 1 / 2-inch drainage type | |
| The 1 group of energy-saving lamps is attached to the observation
window | |
| Equipment shall install four universal casters and fixed casters to
facilitate the movement and fixation of equipment | |
| The equipment is air-cooled | |
- Air supply and temperature measuring system
| Note |
6.1 Air supply characteristics | High uniformity upper wind outlet lower wind return | |
6.2 Cycle motor | Taiwan Shangyu low-pressure asynchronous motor | |
6.3 Wind wheel | Multi-wing centrifugal circulating fan, resistant high and low
temperature aluminum alloy blades | |
6.4 Sensor | PT100 platinum resistance sensor | |
| Note |
7.1 Refrigeration mode | Multiacked mechanical vapor compression | |
7.2 Compressor | 1, One 6P French Taikang compressor, single-stage refrigeration, to
ensure the easy temperature drop to-40℃ | |
7.3 Refrigerant | DuPont non-o HFC R404A/R23 | |
7.4 Condenser | Custom air-cooled, energy-saving condenser | |
7.5 The Heat Exchanger | Swedish Shrep-efficient ST board heat exchanger | |
7.6 The Evaporator | Custom efficient multisegment with hydrophilic fin evaporator (fin
thickening) | |
7.7 Oil separator | Emerson | |
7.8 Expansion valve | Denver thermal expansion valve | |
7.9 Solenoid valve | Heron Palace and Denvers, Japan | |
7.10 Visual mirror | Denvers | |
7.11 Evaporation pressure regulating valve | Denver vaporation pressure regulating valve (guarantee low
temperature and high humidity conditions for long time) | |
7.12 Other accessories | All are international brand quality products (right: dry filter) | |
7.13 Dehumidification | Frozen and dehumidification by refrigeration system (air in the box
passes below the dew point temperature evaporator, water
separation) | |
7.14 System characteristics | The flow of the refrigeration system is automatically switched
under different working conditions, and the excess system is cut
off for energy-saving operation - Cooling control: multi-stage evaporator combination can obtain more
uniform and efficient cooling exchange efficiency and refrigeration
output; flow control is automatically adjusted by their matching
electronic expansion valves;
The 2) control system automatically adjusts the optimal energy
saving condition of refrigerant operation according to test
conditions; condensate pressure regulating valve (cooling water
flow regulating device), evaporative pressure regulating valve
(prevent evaporator freezing), and compressor return cooling
circuit. The controller digital displays the high, low pressure,
exhaust temperature and operating current of the compressor of the
refrigeration system 3) Refrigeration process: the protection measures for the
compressor are fully considered in the design of the refrigeration
system, such as the automatic compressor suction and exhaust
pressure protection function, which keeps the compressor to
maintain the operating temperature within the normal temperature
range, to avoid the compressor too cold or overheating, so as to
prolong the service life of the compressor. High-quality anaerobic
copper pipe gas protection welding method is adopted, which avoids
the damage of oxides on the inner wall of the copper pipe to the
refrigeration system and the compressor. In the design of the
refrigeration system, the vibration reduction measures of the unit
operation, such as the compressor installing spring shock absorber,
and adopting the way of increasing arc bending on the refrigeration
pipeline to avoid pipeline deformation and leakage caused by
operation vibration and temperature changes, thus improving the
reliability of the whole refrigeration system. Other energy saving measures: the following effective energy
regulation measures are adopted, such as refrigeration volume
regulation, gas-liquid bypass regulation, evaporation temperature
regulation, etc. At any low temperature point, the operation power
can be reduced to half, reducing the operation cost and failure
rate of the cooling system to a relatively economic state. 5) compressor air return temperature regulation: automatically
adjust the air return temperature of the compressor to keep the
compressor temperature within the normal range and avoid over-cold
and overheating of the compressor. 6) vibration reduction measures: 1. Compressor: spring vibration
damping.2. Refrigeration system: special rubber pad overall
secondary vibration damping; the refrigeration system pipeline adds
R and elbow to avoid the variation of the copper pipe caused by
vibration and temperature changes, thus causing the rupture of the
refrigeration system pipeline. 7) noise reduction: refrigeration cabinet: wavy special acoustic
sponge. |
- Heating and humidification system
| Note |
| High quality nickel-chrome and gold wire heater | |
| 1) water basin heating and humidification method 2) stainless steel armoured heater 3) heater control mode: contactless and equivalent periodic pulse
broadening, SSR(solid-state relay) 4) water level control, heater dry retardant | |
8.4. Control mode | High precision contact switch control through the SSR solid-state
relay | |
- Controller and control circuits-Taiwan original TT-5166
| Note |
| High precision resistance touch screen, can be controlled with USB
external mouse; | |
| The 7 " TFT true color LCD Chinese and English switch display.
(Resolution: pixel: 800 * 480, temperature ≤ 1℃, time ≤ 1min) |
| 2 Groups of input, support dry and wet ball PT100 and humidity
transmitter, thermocouple T,K,E, etc |
| 0.1S high-precision sampling |
| 120 group program * 1000 segment (1000 cycles respectively), the
number of segments required per group can be arbitrarily divided,
each group program and can freely connect to each other (up to 20
steps) |
| Fixed value (slope) / program operation mode. Change curve meets
the standard requirements |
| The PID + SSR/SCR automatically produces a positive and reverse
bidirectional synchronous output. |
- Software running function
| 1. Temperature and humidity curve: Paper record: built-in recorder with controller memory of 24H for
300 days. It can facilitate the automatic generation of temperature and wet
curve files and transfer into XLS table; 2. complete interface: External USB port: can directly connect the printer, realize local
data printing by line, Easy export of historical curves from a U disk from the USB port. External LAN standard network interface: no special need to
configure the server, you can access the company computer LAN
network, one can connect and monitor up to 16 devices, which is
convenient and fast. With RS-485 communication interface, provide
communication protocol and testing software, and support the
windowsXP and above system with the preservation and printing
function of curve and experimental data.. 3. real-time monitoring can be achieved: simultaneously monitor the
operation status of 1~16 controllers- - - -such as real-time data
of the monitoring controller, signal point status, actual output
status, etc Support for a remote (extranet) fault diagnosis: Through the network, technicians can leave the port, open the IE
browser, enter the IP address of the controller, and enter the
login interface to select the operation or monitor for diagnosis 4. timing control Two sets of timing output control interfaces, combined with 10 time
control modes, can provide the switch signal output according to
the program. Note that only the control signal, the test power
supply in the box should be separately connected with a specially
equipped power supply system. The 5. fault is displayed 16 Failure alarm output, humanized fault cause and troubleshooting
method in Chinese and English The 6. appointment starts All test conditions are available. | |
- Controller protection function
| Overtemperature protection of 1. test products (independent) The 2. comes with a fuse-free protection switch 3. heater overtemperature protection switch 4. compressor overload overheating High and low pressure protection for the 5. compressor 6. system overload over-current protection device, electric leakage
protection 7. water shortage protection function The 8. controller automatically identifies the fault and displays
it (fault alarm and cause, processing prompt) 9. temperature warning and upper and lower temperature
overtemperature The 10. downtime protection function. 11. Calendar timing function (automatic start and automatic stop
operation) 12. POST function The 13. three-tier password protection controller settings
parameters | |
9.10. Control circuit Main components | The main parts of the control line are high quality brands (right
picture: Fuji electromagnetic switch) | |
- Other hardware protection features
| Note |
| 1. studio overtemperature protection, fan overheat protection 2. compressor overpressure The 3. compressor is overcurrent The 4. compressor overheated 5. condensing fan overheating 6. exhaust temperature protection The 7. power supply system 8. power supply phase deficiency, phase order error, underpressure
protection, heater short circuit and other overcurrent protection 9. water supply / drainage abnormal protection 10. humidification pipe overheating protection Earthing protection of the 11. test box housing The 12. comes with a fuse-free protection switch The 13. phase power supply is under the reverse phase protector The 14. emergency stop switch 15. power-off protection function | |
| Note |
11.1 Standard standard | 1.304 Stainless steel test frame type 2, adjustable and free 304
stainless steel put aside 8 (4 on each layer) 2. plastic pull 20L water tank one 3. diameter 50mm test hole, soft plug, 304 stainless steel hole
cover 1 set 4. Operating Manual Maintenance Manual 5. provides a qualified calibration report from a third-party
calibration agency. One set of 6. installation and user manual 7. Equipment Maintenance Manual 1 set 8. 1 set of equipment fault handling manual One set of electrical and gas circuit wiring diagram of 9.
equipment 10. equipment List 1 The 11. software system provides 1 disc backup 12. Certificate of Compliance | |
| Note |
| 1. ambient temperature: 5℃ -35℃; 2. relative humidity: ≤ 85%RH; 3. atmospheric pressure: 80kPa~106kPa No strong vibration around the 4. 5. power supply voltage: 5.1 Connect power supply 380V AC(± 10%), three-phase + protection
ground wire, grounding resistance ≤ 4 Ω; use 3P + N 40A air switch 5.2 Nominal maximum power: approx 15 KW 6. frequency: 50±0.5Hz Water for 7. humidification; resistivity of ≥ 500MΩ;á | |
| Note |
| The production and management of 1. Company operate in strict
accordance with the ISO9001:2000 quality assurance system. Free
training of operators and maintenance personnel for customers
(located in the Company or customer location). 2. complete machine warranty period: one year. The maintenance once
a quarter, a total of 4 times (the province in the factory regular
maintenance, outside the province according to consultation),
lifelong maintenance. 3.'s complete repair service team, providing the most prompt and
professional repair services (repair 2HR reply, 24HR to the plant,
problem exclusion) | |
- Equipment execution and acceptance standards
| Note |
| The 1. is designed and manufactured according to the national
standard of GB/T10586-2006 and GB/T10592-1989.* Verification
standards are according to GB/T5170.2-2008,GB/T5170.5-2008. | Tips: Temperature uniformity of GB/T5170.5-2008. refers to no-load
condition; humidity uniformity also refers to no-load condition! |