KP-1050S Defibrillation Test Pulse Generator for Medical Device Safety

KP-1050S Defibrillation Test Pulse Generator for Medical Device Safety

Voltage measurement accuracy AC 110/220V
Input power AC 110/220V ±10%, 50/60Hz, 425W
Weight 22.5 kg
Brand Name kingpo
Model Number KP-1050S
Certification iso9001 ce
Place of Origin china
Minimum Order Quantity 1
Price 7000
Payment Terms T/T
Detailed Product Description
KP-1050S Defibrillation Test Pulse Generator for Medical Device Safety
Product Overview

The KP-1050S defibrillation test pulse generator evaluates active implantable medical devices' response to defibrillation waveforms, assessing permanent effects, restoration capability, and overall safety compliance with ISO14708-1 and GB 16174.1-2024 20.2 standards.

Key Specifications
Voltage measurement accuracyAC 110/220V
Input powerAC 110/220V ±10%, 50/60Hz, 425W
Weight22.5 kg
Dimensions680 × 460 × 200mm
Test Procedures
Test 1: Damped Sinusoidal Waveform Evaluation

Purpose: Assess permanent effects on devices and reprogrammability after damped sinusoidal defibrillation.

Equipment: Defibrillation test voltage generator producing specific waveform characteristics.

Parameters:

  • Waveform requirements: 1.5ms < Tp < 2.5ms; 3ms < Tw50 < 5.5ms
  • Test voltage Vtest maximum: 140+70V
Test 2: Truncated Exponential Waveform Evaluation

Purpose: Evaluate safety and reprogrammability under truncated exponential defibrillation.

Equipment: Defibrillation pulse generator with specific duration capability.

Arrangement: C1=(150±50)μF, R1=65Ω with coupled switches S1/S2

Parameters:

  • Waveform duration: Td = (10 ± 0.5) ms
  • Test voltage Vtest maximum: 140+70V
Product Features
  • Programmable controller with color touch screen interface
  • Parameter memory retention during power failure
  • Built-in overvoltage protection for stable operation
  • Positive/negative polarity output configuration
  • Single/double phase setting flexibility
  • User-friendly operation with excellent reliability
Technical Specifications
Compliance StandardsISO 14708-1:2014 & GB 16174.1-2024 20.2
Test 1 Voltage140+70V (RLC Circuit)
Damped Sine Waveform1.5ms < Tp < 2.5ms; 3ms < Tw50 < 5.5ms
Test 1 Program3 positive voltage pulses (20±2s interval) followed by negative pulse after 60±2s
Test 2 Voltage140+70V (9.5ms duration)
Test 2 Program3 positive polarity single-phase pulses (20±2s interval) followed by negative pulse after 60±2s
Evaluation CriteriaDevice functionality and reprogrammability post-testing
Test Repetitions1-999 times
Output ModeFloating output