IEC 60601-1-Figure 34 Spark Ignition Test Apparatus
|
11 .2.2 M E EQUIPMENT and ME SYSTEMS used in conjunction with
OXYGEN RICH ENVIRONMENT 11 .2.2.1 RISK of fire in an OXYGEN RICH ENVIRONMENT In ME EQUIPMENT and ME SYSTEMS , the RISK of fire in an OXYGEN RICH ENVIRONMENT shall be reduced as far as possible under NORMAL CONDITION or SINGLE FAULT CONDITIONS (as identified in 11 .2.3). An unacceptable RISK of fire is considered to exist in an OXYGEN RICH ENVIRONMENT when a source of ignition is in contact with ignitable material and there is no means that would limit the spread of a fire. NOTE 1 For oxygen concentrations up to 25 % at one atmosphere or partial pressures up to 27,5 kPa for higher atmospheric pressures, the requirements in 1 3.1 .1 are considered to be sufficient. a) A source of ignition is considered to exist in an OXYGEN RICH
ENVIRONMENT when any of the following conditions exist in
NORMAL CONDITION and SINGLE FAULT CONDITIONS
(including voltage and current): Items 4) and 5) address the worst case where the atmosphere is 1 00
% oxygen, the contact material (for item 5) is solder and the
fuel is cotton. Available fuels and oxygen concentrations
should be taken into consideration when applying these
specific requirements. Where deviations from these worst case
limits are made (based on lower oxygen concentrations or less
flammable fuels) they shall be justified and documented in First, the place(s) within the ME EQUIPMENT where sparking might
cause ignition are identified. Then the material(s) of the
parts between which sparks can occur is identified. Two contact pins made of the material to be considered are placed
in opposition (see Figure 34). One pin has a diameter of 1 mm,
the other of 3 mm. The electrical source is connected to the
pins as shown in Figure 35 to Figure 37. A piece of cotton is
placed close to the contact surfaces of the two pins. The
contacts are constantly flushed by oxygen with a speed of less
than 0,5 m/s via a tube. The cathode is moved to the anode to close
the contacts and pulled back to open them again. A minimum of
300 trials has to be performed before it can be decided that
the sparks do not ignite. If the sparks get smaller because
of bad surfaces of the electrodes, the electrodes are cleaned
with a file. If the cotton gets The situation with the highest voltage or current respectively and no ignition defines the upper limit. A safe upper limit is given by dividing the upper limit of voltage or current respectively with the safety margin factor of three. Figure 34 – Spark ignition test apparatus Figure 35 – Maximum allowable current I as a function of the maximum allowable voltage U measured in a purely resistive circuit in an OXYGEN RICH ENVIRONMENT Figure 36 – Maximum allowable voltage U as a function of the capacitance C measured in a capacitive circuit used in an OXYGEN RICH ENVIRONMENT Figure 37 – Maximum allowable current I as a function of the inductance L measured in an inductive circuit in an OXYGEN RICH ENVIRONMENT TESTING AND MEASURING EQUIPMENT/ALLOWED SUBCONTRACTING IEC 60601-1:2005 + Am.1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance “R” Required “S” May be subcontracted, see OD 2012 “SPTL” Specialized Facility, see IECEE 02-2 “W” Witness testing in the categories “MED” and “MEAS” “3PPS” Three Phase Power Supply required
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Product Tags: IEC 60601-1-Figure 34 3mm Pin Spark Ignition Test Apparatus 1mm Pin Spark Ignition Test Apparatus |
![]() |
ISO 5364 -Annex C Medical Test Equipment ISO 5364 Test method for patency of lumen |
![]() |
ISO 5364 Apparatus For Testing Resistance To Collapse Of The Buccal Portion |
![]() |
IEC 61010-2-012-Figure 103 Scratching Tool Hard-soldered carbide tip K10 Material |
![]() |
Electrosurgery Analyzer Essential Medical Test Equipment for Accurate and Power Voltage and Current Analysis |
![]() |
Medical Test Equipment for IEC Residual Voltage Testing Includes Cables and Calibration Certificate |
![]() |
IEC 60601-2-2 Testing device for neutral electrodes |