China Lab Testing Equipment manufacturer
SKYLINE INSTRUMENTS CO.,LTD
Skyline Instruments Co.,LTD We only manufacture Lab Testing Equipment
9
Home > Products > Electronic Testing Equipment >

Cable Electronic Testing Equipment IEC60695 Needle Flame Test Chamber

Cable Electronic Testing Equipment IEC60695 Needle Flame Test Chamber

Brand Name SKYLINE
Model Number SL-SE03
Certification CE
Place of Origin China
Minimum Order Quantity 1 unit
Price Negoitable
Payment Terms Western Union, MoneyGram, T/T
Supply Ability 1 pcs/ month
Delivery Time 15 working days
Packaging Details plywood case
SKYLINE Needle Flame Test Chamber
Nner hole diameter inner hole diameter Φ 0.5mm ± 0.1mm, Diameter,9mm≤Φ,
Burner angle 0,°20°, 45° (adjust flame height)
Power 800VA ,220V, 48-60Hz
Chamber ≥0.5m3
Standards IEC884-1,VDE, IEC335,UL1694
Detailed Product Description

Cable Electronic Testing Equipment IEC60695 Needle Flame Test Chamber

 

Introductions

Needle Flame Test Chamber is used to test the resistance of fire caused by an interior fault such as leakage current in cable & wire and the like. It meets the requirements of Chinese and IEC60695-2-2 and IEC61347-1 standards.

 

Application

Electrician & electron product
Components and accessories of household appliances


Features
a.All stainless steel and brass constructions.
b.Burner can be ignited automatically.
c.Temperature and test time display by digital meters.
d.A big window in the door makes your viewing more convenient.

 

Technical parameter

 Needle burner

 inner hole diameter Φ 0.5mm ± 0.1mm, Diameter,9mm≤Φ,

length ≥ 35mm, stainless steel

 Burner angle 0,°20°, 45° (adjust flame height)
 Thermocouple

 Ø0.5mm Type K (Ni/Cr - Ni/Al), armored sheath can withstand

temperature-40°C~ 1100°C

 Burning time 0-999.9 s ± 0.1 s adjustable (usually selected for 30s)
 Hold Burning time 0-999.9 s ± 0.1 s, automatic recording, manual time-out
 Flame Height 12mm ± 1mm (with height measurement instruments)
 Flame Gas butane gas/Butane gas, min. 95% purity (can use bottled lighter gases)
 Temperature Test Range 0~1000°C
 Flame Temperature Requirements from 100 °C ± 2 °C liters to 700 °C ± 3 °C time in 23.5 seconds ± 1 seconds
 Standard Copper Block 4mm±0.01mm, weight.0.58g±0.01g before drilling. Cu-ETP
 Test process automatic control, independent convulsions
 Chamber ≥0.5m3
 Control Using single chip microcomputer + touch screen control
 Consult standard GB/T5169.10-2006 , IEC60695-11-5:2000,UL746A
 Power Supply 800VA ,220V, 48-60Hz
 Display panel language English(customize available)
 
Product Tags: sharp edge tester   flame test chamber  
Related Products
Email to this supplier
 
From:
Enter your Email please.
To: SKYLINE INSTRUMENTS CO.,LTD
Subject:
Message:
Characters Remaining: (80/3000)