Pool 4 Digital LCD 265mV Redox Waterproof ORP Meter
|
|
High Precision ORP/Redox Tester Waterproof ORP Meter Tester Potential Positive and Negative
Operation: Remove the prrotective cap. Turn the meter on by pressing the ON key. immerse the electrode in the solution to be tested. Stir gently and wait for the reading to stabilize. Press the "HOLD" key, the reading is be kept. when repeat measuring, please press the "HOLD" key again, it automatacally relieves the hold condition. After use, turn the meter on by pressing the OFF key. rinse the electrode with distilled water to minimize contamination. replace the protective cap after use.
notes: it is waterproof design, with replaceable electrode.
Calibration: Immerse the meter in the distilled water. Activate it after a few hours. Immerse the meter in the calibration solution. and stir it gently. Regulate the trimmer with a screwdriver until LCD screen for 265mV. Rinse the electrode with distilled water, and suck it with filter paper.
Technical Specification: Measuring range:0~±1999mV Accuracy: ±5mV Resolution:1 mV Display:4-digital LCD Power Supply:4×1.5V(AG13) Dimensions:188×35×35(mm3) Weight:118g |
| Product Tags: Redox Waterproof ORP Meter 265mV Waterproof ORP Meter Pool ORP Redox Meter |
|
Yieryi Digital ORP Meter with LCD Backlight Compact Pen Size and CE RoHS Marking for Water Quality Testing |
|
501-2 ORP Electrode with Ceramic Core Ag/AgCl Reference and ±2000 mV Range for Lab and Field Use |
|
1999mV Range ORP Meter with LCD Backlight and Portable Pen-Type Design for Water Quality Testing |
|
Platinum Needle ORP Probe with BNC Q9 Connector for 0~+1500mv Range Oxidation-Reduction Potential Testing |
|
ORP Electrode Sensor with ±5mV Accuracy, 0~1500mV Range, and IP65 Protection for Wastewater & Lab Use |
|
Waterproof ORP Meter with Replaceable Electrode and 0~±1999mV Range for Accurate Redox Testing |