China Rubber Testing Machine manufacturer
Dongguan Zhongli Instrument Technology Co., Ltd.
Providing one-stop service for global laboratories, customer's satisfaction is our ultimate goal!
9
Home > Products > Environmental Test Chamber >

UL1581 ASTM D2436 Air Ventilation Aging Testing Chamber

UL1581 ASTM D2436 Air Ventilation Aging Testing Chamber

Brand Name Zhongli
Model Number ZL-6010A
Certification ISO,CE
Place of Origin Guangdong,China
Minimum Order Quantity 1 set
Price Negotiable
Payment Terms T/T,L/C,D/A,D/P,Western Union,MoneyGram
Supply Ability 100 sets per month
Delivery Time 7~10 days after the order confirmed
Packaging Details UL1581 ASTM D2436 Air Ventilation Aging Testing Chamber Packaged in wooden case
Product name UL1581 ASTM D2436 Air Ventilation Aging Testing Chamber
Power supply Three-phase 380VAC 50Hz
Ambient temperature 5~35℃
Ambient humidity ≤85% R.H
Temperature range Ambient temperature +25℃~300℃
Control accuracy ±0.5℃
Resolution accuracy 0.1℃
Distribution error ±2.5℃
Detailed Product Description
UL1581 ASTM D2436 Air Ventilation Aging Testing Chamber
Professional testing chamber designed for air ventilation aging tests according to UL1581 and ASTM D2436 standards, ensuring precise environmental simulation for material testing applications.
Performance Specifications
ParameterSpecification
Power SupplyAC380V
Power6KW
Temperature RangeAmbient temperature +25℃~300℃
Control Accuracy±0.5℃
Resolution Accuracy0.1℃
Distribution Error±2.5℃
Heating TimeFrom ambient temperature to 300℃, approximately 60 minutes
Machine Structure & Materials
  • Inner chamber dimensions: 45cm (depth) * 50cm (width) * 50cm (height)
  • Outer dimensions approximately: 1040mm (length) * 650mm (width) * 1275mm (height)
  • Inner chamber material: SUS#304 stainless steel, outer case: baked paint
  • Insulation material: High-density rock wool
  • Heater: Spiral tube heater set: AC380V 6KW
Air Exchange Rate Test Procedure
Comprehensive testing methodology for determining air exchange performance in aging test chambers:
  • Seal all ventilation holes, doors, thermometer holes, and gaps around the blower using pressure-sensitive tape
  • Connect a standard watt-hour meter to the aging chamber's power system
  • Control working chamber temperature to 80±2℃ above ambient temperature
  • Measure electrical energy consumption after temperature stabilization for 2 hours
  • Remove seals and repeat measurement with air exchange port open
  • Calculate air exchange rate using standardized formula
Air Exchange Rate Calculation
N = 3590 • (P2 - P1) / V • d • △T
Where:
  • N = Air exchange rate (times/hour)
  • P1 = Average electrical power consumed when sealed (W)
  • P2 = Average electrical power consumed when open (W)
  • △T = Difference between chamber temperature and ambient temperature (℃)
  • V = Working chamber volume (L)
  • d = Density of air at surrounding ambient temperature (g/L)
Forced convection types: Up to 100 times/hour or 100-200 times/hour (adjustable)