China Fluid Level Meter manufacturer
Xi'an Kacise Optronics Co.,Ltd.
Kacise, understanding your automation needs.
7
Home > Products > Water Quality Sensor >

KEC601 0.6MPa Water Quality Monitoring Sensors 3/4 NPT Thread

KEC601 0.6MPa Water Quality Monitoring Sensors 3/4 NPT Thread

Brand Name kacise
Model Number KCD602
Certification CE
Place of Origin CHINA
Minimum Order Quantity 0-100
Price $0-$1500
Payment Terms D/A, L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability 100
Delivery Time 3-10days
Packaging Details Common package or custom package
Measuring range 0~5000μS/cm
water quality monitor Conductivity Resolution 1μS/cm
water quality monitor Conductivity Precision ±1.5%F.S.
Working Temperature 0~65℃
Power Supply 12~24VDC ±10%
Signal Output RS-485(Modbus/RTU)
Wetted Material ABS/SUS316L(Default)/PVC customized
Detailed Product Description

KEC601 Small Outline Online Water Quality Monitor Conductivity Sensor For Drinking Water Surface Water Various Water

1.Application Environment

  • Drinking water / surface water / various water supply / industrial water treatment.
  • Signal output: RS-485 (Modbus/RTU protocol).
  • Convenient connection to third-party equipment such as PLC, DCS, industrial control computer, general controller, paperless recording instrument or touch screen.
  • Immersion installation with 3/4 NPT pipe thread for easy submersible installation or installation in pipes and tanks.
  • IP68 protection grade.

2.Technical Specifications

ModelKEC601
Measuring range0~5000μS/cm
Resolution1μS/cm
Precision±1.5%F.S.
Working Temperature0~65℃
Working Pressure<0.6MPa
Power Supply12~24VDC ±10%
Signal OutputRS-485(Modbus/RTU)
Wetted MaterialABS/SUS316L(Default)/PVC
Mounting MethodImmersion mounting, 3/4 NPT thread
Cable Length5 meters, other lengths can be customized
Temperature CompensationAuto temperature compensation(Pt1000)
CalibrationTwo-point calibration
Power Consumption<0.3W@12V
Protection GradeIP68
 

3.Dimensional Drawing

 

4.Installation and Electrical Connection

4.1 Installation

Note: At least 2 cm from the bottom and side walls of the container during installing and testing.

4.2 Electrical Connection

a) Red line - power cord (12 ~ 24V)

b) Black line - ground (GND)

c) Blue line - 485A

d) white line - 485B

e) bare wire - shielded wire

After wiring is completed, it should be carefully checked to avoid incorrect connections before powering up.

Cable specification: Considering that the cable is immersed in water (including sea water) for a long time or exposed to the air, the cable has certain corrosion resistance. The outer diameter of the cable is Φ6 mm and all interfaces are waterproof.

 

 

 

 

 

 

Product Tags: 0.6MPa water quality monitoring sensors   3/4 NPT thread water monitoring sensors   24VDC water quality monitoring sensors  