16KPa Pressure 2mL Synthetic Blood Penetration Tester Time Accuracy 0.1s
|
|
Medical Masks 16KPa Pressure 2mL Synthetic Blood Penetration Tester
Product overview:
Test method:
The medical mask sample is supported on the test device, and 2mL of synthetic blood with a surface tension of 0.042 ± 0.002n / m is sprayed from a cannula with an inner diameter of 0.84mm at a distance of 300mm to the mask under test at a pressure of 16kpa in the horizontal direction. Simulate the scene of the mask being splashed by blood from the perforated blood vessel. Take off and visually check the penetration of synthetic blood on the other side of the mask within 10 seconds. Test the medical mask with the corresponding liquid jet pressure of 450cm / s, 550cm / s, 635cm / s, record the test result of each jet velocity, give out the highest blood pressure corresponding to the acceptable quality level limit of 4 for medical masks.
Technical parameters:
|
|||||||||||||||||||||
| Product Tags: mask testing machine mask testing equipment | |||||||||||||||||||||
|