China Electrical Appliance Testing Equipment manufacturer
Sinuo Testing Equipment Co. , Limited
Sinuo Testing Equipment Co., Limited
7
Home > Products > Flammability Test Chamber >

Cellular Plastic Materials Flammability Test Chamber Horizontal Burning PLC Control ISO9772

Cellular Plastic Materials Flammability Test Chamber Horizontal Burning PLC Control ISO9772

Brand Name Sinuo
Model Number SN777
Certification Calibration Certificate (Cost Additional)
Place of Origin China
Minimum Order Quantity 1
Price Negotiable
Payment Terms T/T, L/C, Western Union, MoneyGram
Supply Ability 10 Sets Per Month
Delivery Time 20 Days
Packaging Details Plywood
Control Mitsubishi PLC+ Remote Controller
Operation 7" WEINVIEW touch screen
Gas Pressure 0.1MPa
Gas Flow 965±30ml/min
Flame Height 38±2mm
Test Volume >0.5 m³
Detailed Product Description

Foam Horizontal Burning Test Chamber ISO9772 Cellular Plastics Small Flame Testing Apparatus

 

Overview:

 

This test chamber adopts PLC control and 7" touch screen display, the operation is Intelligent and humanized; uses the integral intake and automatic ignition system, the combustion time delays 0.1S thus to make sure the sufficient time of gas burning; equipped with imported brand flowmeter to ensure the accuracy of the testing.          

 

 

                                    

Principle:

 

This horizontal burning test equipment is intended to process a small-scale laboratory screening procedure for comparing the relative burning characteristics of horizontally oriented, small cellular plastic specimens having a density less than 250kg/m-³ determined in accordance with ISO 845, when exposed to an ignition source. It meets the requirements of  ISO9772-2012, UL94.

 

Technical Elements:

 

Working voltage230V/50Hz or 110V/60 Hz (can be customized)
ControlMitsubishi PLC+ Remote Controller
Operation7" WEINVIEW touch screen
P/PF2 laboratory burnerISO10093,inner diameter Φ9.5±0.3mm,having a barrel  length:100±10mm
Burner wing top

Having an opening of internal length (48±1)mm and internal width (1.3 ±0.05)mm


Gas pressure0.1MPa
Gas flow and gas category965±30ml/min, for the methane gas purity of 98% , it can also use the mixed gas or propane gas, flow meter adjustment
Combustion / self-extinguishing time0-9999X0.1S
Distance between the wing top of the blast burner to the test sample

13±1mm


Flame height38±2mm,blue flame
Support net215×75×13mm, grid: Φ0.8 made of stainless-steel wire with length of the side of 6.4mm
LightingEquip with lighting in the test chamber, with control
Test chamber and dimension>0.5 cube,black matte background,with exhaust fan,dimension:1100*550*1200mm,130kg
Testing environment

Temperature: 15°C~35°C

Humidity: 45%~75%

 
 
How to carry out the testing procedures on this chamber?
 

1) Press  Start button for 3-5s or button 1 on the remote control to light up the flame. Adjust the flame height according to the standard requirements.

2) Close the window door and re-posit the burner under the specimen according to standard requirements, press Start Timing button or button 3 on the remote control to start the flame application time timing. The flame application is 60s, and after reaches 60s the flame is extinguished, and the Elapsed time te starts timing.

3) Press te Pause button or button 3 on the remote control to stop the Elapsed time te timing when the specimen flam reaches the 60mm gauge mark.

4) Press te/tb Pause button or button 5 on the remote control to stop the Elapsed time te and Burning time tb timing when the specimen ceases to burn or glow before reaching 60mm gauge mark.

5) Press [tb Pause / Start] to be tb Start or button 4 on the remote control to start the Burning time tb timing when the specimen flam reaches the 25mm gauge mark, whether the burning is on the bottom, top or edge of the specimen.

6) Press [tb Pause / Start] to be tb Pause button or button 4 on the remote control to stop the Burning time tb timing when the flam or glowing combustion front reaches the 125mm gauge mark, or when the specimen ceases to burn or glow before reaching 125mm gauge mark.

7) Operator should calculate out the distance burnt (Ld): this is the distance between the 25mm gauge mark and the point where the flame or glowing combustion front stopped, expressed in millimetres. If the flame front went out before the 25mm mark, record that Ld =0.

8) After each test, open the exhaust fan in time and then clean the support gauze, burner, wing top and chamber; if operator needs to process the second test, press [Count Cleared] to clear all the timers firstly, allow the support gauze to cool to room temperature, check the flame at least once every five tests.

 


Product Tags: flame test chamber   flammability test apparatus  