AC Hipot test |
Booster rated output capacity | 500VA (5000V/100mA), short circuit current is greater than 200mA |
Output voltage waveform, distortion, adjustment rate | Sine wave, <3% (pure resistance,5000V/1mA), <3% (from empty
load to full load) |
Output voltage setting | Range | 500~5000V |
Resolution, accuracy | 5V,±(3%×setting+10V) |
Frequency setting | Range, accuracy | 50Hz/60Hz, ±0.1%×setting |
Alarming limits setting | Current upper-limit | 0.10~99.99mA |
Current lower-limit | 0.00~10.00mA |
Compensation current setting | Range | 0.00~10.00mA, automatic testing, compensation can be set to ON/OFF |
Time setting | Range | 1~999s |
Ramp up time range | 1~100s |
Accuracy | ±(0.1%×setting+1count) |
Breakdown current alarm | Range, accuracy | 0.10~99.99mA, ±(3%×reading+3counts) |
Insulation Resistance test |
Rated output capacity | 2VA(1000V/2mA) |
Output voltage regulation, output voltage ripple | <3%(From no load to full load), <1% |
Output voltage setting | Range, accuracy | DC 500V or 1000V, ±(3%×setting+3counts) |
Alarming limits setting | Resistance upper-limit range | 0.0~99.9 MΩ,100~2000MΩ(If upper-limit is set to 0, it will not make
adjudgment for upper limit) |
Resistance lower-limit range | 0.3~99.9 MΩ, 100~2000MΩ |
Time setting | Range, accuracy | 1~999s, ±(0.1%×setting+1count) |
Insulation resistance measurement | Range | 0.0~99.9 MΩ, 100~2000 MΩ |
Accuracy | ≤200MΩ: ±(3%×reading+3counts) >200MΩ:±(8%×reading+8counts) |
Ground Bond Resistance test |
No load output voltage | <12V |
Output current setting | Range, accuracy | 3~30A, ±(3%×setting+0.2A) |
Current frequency setting | Range, accuracy | Sine waveform 50Hz/60Hz, ±0.1%×setting |
Test time setting | Range | 1~999s |
Accuracy | ±(0.1%×setting+1count) |
Result display mode | Ground resistance or voltage |
Ground resistance display mode |
Alarming limits setting | Resistance upper-limit range | 3A ≤ output current ≤ 10A:10~600 mΩ 11A ≤ output current ≤ 25A:10~300 mΩ 26A ≤ output current ≤ 30A:10~200 mΩ |
Resistance lower-limit range | 0~100 mΩ |
Compensation resistance setting | Range | 0~200 mΩ, automatic testing, compensation can be set to ON/OFF |
Ground bond resistance measurement | Range, accuracy | 10~600 mΩ, ±(3%×reading+ 5counts) |
Ground voltage display mode |
Alarming limits setting | Voltage upper-limit range | 30~7500 mV |
Voltage lower-limit range | 0~3000 mV |
Compensation voltage setting | Range | 0~1000 mV, automatic testing, compensation can be set to ON/OFF |
Ground bond resistance measurement | Range, accuracy | 10~7500 mV, ±(3%×reading+ 50mV) |
Leakage Current test |
Leakage current test type setting | Single phase load, static/dynamic leakage current |
MD network selection | Standard IEC 60990 figure 4 network; other test network optional |
Alarming limits Setting | Current upper-limit range | 0.050~9.999mA |
Current lower-limit range | 0.000~5.000mA |
Leakage current compensation setting | Range | 0.000~1.000mA, automatic testing, compensation can be set to ON/OFF |
Time test setting | Range, accuracy | 1~999s, ±(0.1%×setting+1count) |
Leakage current measurement | Range | 0.050~9.999mA |
Accuracy | DC~10kHz: ± (3%×reading+0.010mA) 10 kHz ~1MHz: ± (5%×reading+0.050mA) |
Output voltage measurement | Range, accuracy | 60~280V(phase voltage), ±(0.5%×reading+2counts) |
Run Power Only forAN9651S(F)/AN9651F(F)/AN9651B(F) |
Alarming limits setting | Power upper-limit range | 6~6000W |
Power lower-limit range | 0~6000W |
Test time setting | Range, accuracy | 1~999s, ±(0.1%×setting+1count) |
Voltage measurement | Range, accuracy | 60~280V (phase voltage), ±(0.5%×reading+2counts) |
Current measurement | Range, accuracy | 0.030~3.999A, 4.00~25.00A, ±(0.5%×reading+2counts) |
Active power measurement | Range, accuracy | 30.0~199.9W: ±(0.5%×reading+5W), 200~6000W(or2000W): ±(0.5%×reading+30W) |
Power factor measurement | Range, accuracy | 0.10~1.00, ±(2%×reading+2 counts) |
Low voltage start test |
Alarming limits setting | Current upper-limit range | 0.30~25.00A |
Current lower-limit range | 0.00~25.00A |
Test time setting | Range, accuracy | 1~999s, ±(0.1%×setting +0.05s) |
Voltage measurement | Range, accuracy | 60~280V (phase voltage), ±(0.5%×reading+2counts) |
Current measurement | Range, accuracy | 0.030~3.999A, 4.00~25.00A, ±(0.5%×reading+2counts) |
Power supply for UUT |
SN2212B Built-in isolation transformer |
Isolation transformer input voltage, frequency | Determined based on input voltage and frequency |
Isolation transformer output voltage ratio | Leakage current test | The output voltage is 1.06 times the input voltage |
Run power test | The output voltage is 1.00 times the input voltage |
Low voltage start test | The output voltage is 0.85 times the input voltage |
Isolation transformer capacity | Single phase load rated capacity 6000VA |
Isolation transformer voltage regulation | 3% (from no load to full load) |
Overcurrent protection | Maximum current is 30A, automatically cut off the isolated power
supply after 5s overcurrent |
SN2212S, SN2212F Built-in variable frequency power supply |
Max. output capacity | 5000VA |
Output voltage distortion | 1% @ 47~63Hz |
Output voltage stability | 1% |
Programmed output voltage | Voltage range | 0~300V |
Voltage accuracy | 0.5%RD+0.5%F.S. |
Programmed output frequency | Frequency Range | 47~63Hz |
Frequency stability | 0.1% |
Output maximum current | 0~150V file | 27A |
0~300V file | 14A |
SN2212S Configured IN-Pad |
Processor | ARM926 core, 400MHz frequency |
Operating system | Windows CE |
Software | ESRS-Pad special software for safety test |
Display | 10.4” TFT color LCD |
Operation method | Input through touch screen, support input through USB interface
mouse and keyboard |
Storage | 50MB FLASH, user can choose large capacity SD memory card |
External Interface | 2-way RS232 interface, 2-way USB interface, CAN interface, DC12V
power interface |
Mounting brackets | 3D, full angle, display stand |
Bar code scanner | One barcode scanner with an RS232 interface defaulted |
Interface function |
Remote control interface | Standard, standard foot switch (start) |
Alarm light interface | Standard, standard three-color warning light (test, qualified,
warning) |
Communication Interface | Standard RSA32, optional RS485, network port, GPIB, etc. |
Variable frequency power control interface | Optional |
Switch interface | Optional, start, stop, test, pass, alarm switch output |
Printer interface | Optional, support for dot matrix printers and micro printers |
General specification |
Verification and calibration environment | 23 ° C ± 5 ° C, (45 ~ 75)% RH, less dust |
Use environment | 0 ° C ± 40 ° C, (5 ~ 90)% RH, less dust |
Storage environment | -10 ° C ~ 55 ° C, (5 ~ 95)% RH, less dust |
System power | 220V±10%, 50Hz±5%, 10A |
System power consumption | Maximum 1000VA (excluding the operating power consumption of the
tested product) |
System weight (without packaging) | About 80kg |
Dimensions (mm, without packaging) | SN2212S | 483(W)x 600(D)x 1355(H) |
SN2212F | 483(W)x 600(D)x 1355(H) |
SN2212B | 530(W)x 460(D)x 1105(H) |