| Dimensions (H x W x D) | 179 x77 x 42 mm |
| Weight | Approx. 495 g, depending on sensor configuration, without strap,
without pump Approx. 550 g, depending on sensor configuration, without strap,
with pump |
| Housing | Durable two-component housing |
| Display | High-contrast colour display |
| Temperature | -20 ℃ to 50 ℃ |
| Pressure | 700 to 1,300 hPa (measuring function) 800 to 1,100 hPa (use in explosion-hazard areas) |
| Relative humidity | 10 to 90 % (short-term up to 95 %)r.h. |
| Alarms | Visual: | 3 LED 'red'(gas alarms), 3 LED 'yellow'(device alarms) |
| Acoustic | Multi-tone, typically 100 dB(A) at 30 cm |
| Vibration | |
| Ingress protection class | IP 68 |
| Energy supply | Lithium-ion battery, rechargeable, inductive charging |
| Operating times (Diffusion) | With CatEx and 3 EC sensors | Typically 24 hours |
| With lR and 3 EC sensors | Typically 22 hours |
| With 3 EC sensors | Typically 120 hours |
| With CatEx, PlD and 3 EC sensors | Typically 17 hours |
| With lR, PlD and 3 EC sensors | Typically 16 hours |
| With CatEx-, IR- and 3 EC sensors | Typically 14 hours |
| PID only | Typically 42 hours |
| Charging times | Typically 4 hours after use during a shift of max. 10 hours |
| Start-up times | Typically <60 seconds for standard sensors |
| Data storage | 24 MB, e.g. at 10 minutes per hour of gas exposure with measuring
values changing bythe second on all 7 channels: approx. 400 hours |
| Pump operation | Max. hose length 45 m |
| Approvals | Marking Explosion Protection: |
| ATEX /IECEX | I M1, II 1G |
| Ex da ia l Ma, Ex da ia llC T4 Ga |
| EAC | PO Ex da ia l Ma X |
| OEx da ia llC T4 Ga X |
| cCSAus | Class l, Zone 0, AEx da ia llC T4 Ga |
| Class ll, Div 1, Gr. E,F,G |
| C22.2 No.152,ANSISA 12.13.01:2000 |
| ANZEx | Ex da ia l Ma, Ex da ia llC T4 Ga |
| Measurement Performance: | Explosion protection according ATEX: EN 60079-29-1(CatEx 125 PR, CatEx125 PR Gas, (Dual) IR Ex, XXS H2
HC) CatEx 125 PR and (Dual) IR Ex: methane,propane, n-nonane, etc. The following applies to the CatEx 125 PR and approved gas
"n-nonane": Only in conjunction with pump adapter "Nonan"part
number: 3720225 |
Oxygen deficieny/excess oxygen: EN 50104 (XXS O,/ XXS O,PR) |
Toxic gases: EN 45544-1, EN 45544-2,EN 45544-3(XXS H,S LC, XXS CO LC, (Dual) IR
COz) |
| Software: EN 50271 |
| Directives | 2014/34/EU(ATEX) |
| 2014/30/EU(EMV) |
| 2011/65/EU(RoHS) |
| 2014/90/EU (MED) |
| Marine Approval | DNV GL |
| Manufacturer´s warranty | 3 years for the device |
| 1 year for the power supply |
| Sensors: see DrägerSensor"& Portable Instruments Handbook |