Simens X700 Heart Ultrasound Machine Repair High Frequency Linear Transducer
|
Simens X700 heart ultrasound machine RepairSimens X700 System Reports Error
Fault Phenomenon
The self-check before the device is used indicates that the flow sensor test fails.
Fault analysis and Troubleshooting
Analyze the factors that cause the flow sensor test failure
1.Insufficient air source pressure does not meet the test standard; 2.The outgoing call box is faulty; 3.The gas module is faulty. 4.Internal circuit boards of the ventilator, such as PC1772 and PC1771, are faulty.
Check the air source pressure is normal, eliminate the problem of air source. Further check the outgoing box. After replacement, the power-on self-test passes, indicating that the fault is in the outgoing box. Siemens ventilator is an integrated exhalation box, which allows high-temperature disinfection. However, high-temperature and high-pressure sterilization equipment in domestic hospitals cannot meet the sterilization level of exhalation box, so the high-temperature and high-pressure sterilization is basically abandoned in China. If the disinfection method is not correct, it will often cause the failure of the exhalation box. The key points of cleaning and disinfecting exhalation box are as follows: First, do a good job of cleaning and disinfecting the exhalation box before sending it to the supply room for sterilization. The tubes of the exhalation box will be stained with the patient's body fluid, so the tubes of the exhalation box can be washed clean by cleaning and disinfection. The common method is to soak the tubes with diluted hypochlorous acid disinfectant, and then send them to the supply room for sterilization after cleaning and disinfection.
Secondly, after the cleaning, disinfection and sterilization work in the supply room, the exhalation box must be dried. Only the use of the dried exhalation box can ensure safety, otherwise it is very easy to damage the exhalation box.
Ultrasound parts that we can repair:
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Product Tags: Simens X700 Heart Ultrasound Machine Repair Linear Transducer Ultrasound Machine Repair |
![]() |
Invalid Ultrasound Machine Repair Hitachi ARIETTA 60 Occasional Hints That Probe Detection |
![]() |
HD11 XE Freeze Key Not Sensitive Ultrasound Machine Repair Monitor |
![]() |
Repair CX50 Ultrasonic Mainboard 453561368033 453561622163 |
![]() |
Aloka Hivision Ascendus Ultrasound Machine Repair Solutions Cell Board |
![]() |
HD15 ACB Board Medical Ultrasound Machine Repair Solutions |
![]() |
Hitachi Aloka Cell Board For Arietta 70 Ultrasound System Repair Solutions |