Preirus Troubleshoot Ultrasound Machine Repair Hitachi Aloka
|
Hitachi Aloka HI VISION Preirus troubleshoot color image interference
Fault Phenomenon The color ultrasound diagnostic apparatus has a hot wheel
interference during use, which seriously affects the diagnosis of
the image. Fault analysis and Troubleshooting The color ultrasound diagnostic apparatus has a hot wheel interference during use, which seriously affects the diagnosis of the image. According to the fault phenomenon, each probe of the machine in use has the same interference, and the possibility of failure of the color ultrasonic diagnostic apparatus is preliminarily ruled out. According to the safe use specifications of the color ultrasonic diagnostic equipment, the grounding end of the machine shell and the machine power plug should be grounded. The loop formed by improper grounding can also cause electromagnetic interference; then, use a power measuring instrument to measure the grounding resistance. 2 ~ 5Ω, but after checking that the machine is not installed with a ground wire; after reinstalling the ground wire, the resistance value of the test ground wire is 3Ω, which meets the standard; the power-on test, the interference disappears, and the fault is eliminated.
Ultrasound parts that we can repair:
The solution of dealing with ultrasonic image interference
1. Probe interference: replace the probe 2. Equipment interference: replace the power supply or internal boards 3. External interference: (1) Workstation interference: it is recommended to add several magnetic rings to the video cable, if it is not work, replace the ATX power supply, or add a filter to the host power input. (2) Power input interference: add several magnetic rings to the power line.
If the above methods do not work, just try the method of eliminating the equipotentiality , adding a shielding cloth on the bed, and connect a wire to the ground strap of the machine. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Product Tags: Preirus Troubleshoot Ultrasound Machine Repair Hitachi Aloka Ultrasound Machine Repair |
![]() |
Invalid Ultrasound Machine Repair Hitachi ARIETTA 60 Occasional Hints That Probe Detection |
![]() |
HD11 XE Freeze Key Not Sensitive Ultrasound Machine Repair Monitor |
![]() |
Repair CX50 Ultrasonic Mainboard 453561368033 453561622163 |
![]() |
Aloka Hivision Ascendus Ultrasound Machine Repair Solutions Cell Board |
![]() |
HD15 ACB Board Medical Ultrasound Machine Repair Solutions |
![]() |
Hitachi Aloka Cell Board For Arietta 70 Ultrasound System Repair Solutions |