Mindray M8 M9 TE7 Ultrasound SV300 Ventilator Rechargeable Li-ion Battery LI24I002A
|
|
Mindray M8 M9 TE7 Ultrasound SV300 Ventilator Rechargeable Li-ion Battery LI24I002A
► Brief Introduction The ventilator is designed to operate on battery power whenever power supply becomes interrupted. When the ventilator is connected to the external power source, the batteries are charged regardless of whether the ventilator is currently on or not. In case of power failure, the ventilator will automatically be powered by the internal batteries. When external power source is restored within the specified time, power supply is switched from battery to external power supply automatically to ensure continuous system use.
► Battery Performance Checking Check battery performance once every six months. Checking battery performance is also required before ventilator repair is carried out or when battery is doubted to be the source for ventilator failure. Battery performance may degrade over time. Follow these steps to check battery performance: 1. Disconnect the patient from the ventilator and shut down the ventilator. 2. Connect the ventilator to the external power source and charge the batteries uninterruptedly for at least 10 hours. 3. Disconnect the external power source. Allow the ventilator to operate on battery power until the ventilator shuts off. 4. The running time of the battery reflects its performance. If the running time of the battery is noticeably shorter than that stated in the specifications, replace the battery or contact the service personnel.
► Battery in Use 1. Connect the ventilator to AC power and push the hardkey to switch on. 2. Disconnect the AC power after the system starts up. 3. Verify that the [Battery in Use] alarm is activated and the ventilator is powered by batteries. 4. Reconnect the AC power. 5. Verify that the alarm resets and the ventilator is again powered by AC.
► Remove the Lithium Battery
1. Lay down the machine. Remove the 4 M3X8 self-locking screws on the main unit battery compartment door from the bottom to remove the battery compartment door.
2. Pull out the slice on the battery to remove the lithium battery.
► Specification
► Shipment Arrangements As for delivery time, we will try our best to process the orders and ship them out as soon as possible. Please be patient and keep contact with us. Generally speaking, most orders will be shipped in 3 to 5 working days and can arrive on time. However, during the busy days the shipment may be delayed over 5 working days. If you are in urgent need of your item, please be sure to confirm with us before placing your orders. To make sure the business goes smoothly, please pay special attention to it and carefully arrange your time of placing orders according to your personal needs. If your order is taking longer than expected or you want to make it clear that when it will arrive, please feel free to contact us. We are always here to help you out.
"Quality assurance" is our eternal pursuit and promise. Don't be hesitant to contact us! Welcome to place your orders. Thank you for your time and interest in Mindray M8 M9 TE7 Ultrasound SV300 Ventilator Rechargeable Li-ion Battery LI24I002A .
|
||||||||||
| Product Tags: LI24I002A Rechargeable Li-ion Battery SV300 Ventilator Rechargeable Li-ion Battery |
|
FUJIFILM 125N120008 Rechargeable Li-Ion Battery 7.4V 3200mAh 24Wh New Original for DR1200 DR1270 |
|
Bellavista 1000 Annual Maintenance Service Kit PN 301.972.000 New Original Filter Medical Accessories |
|
GE 03-50049-001 Patient Monitor Accessories Rechargeable Lithium Battery |
|
IMEC5 IMEC8 IMEC10 IMEC12 Patient Monitor Li-Ion Battery LI131001A Mindray Ipm5 Ipm8 Ipm10 Ipm12 |
|
GE MAC3500 Battery 900770-001 New Original MAC PAC 18V 3500mAH GEMS-IT Medical Accessories |
|
GE 1505-3231-000 Flow Sensor TRANSDUCER FLOW 0-200 L MIN Original LOT CF0222 |