SonoSite HFL38X 13-6 MHZ Ultrasound Transducer REF P07682-20A
|
|
SonoSite HFL38X 13-6 MHZ transducer REF P07682-20A
SonoSite HFL38X 13-6 MHZ transducer REF P07682-20ASpecification
Cleaning and disinfection of the transducer (cleaning method)
1. Disconnect the drive from the system.
2. Remove all parts of the sensor housing.
3. Wipe the surface with a soft cloth lightly moistened with a mild soap or cleaning solution to remove any particles or body fluids. Apply the
solution to the cloth, not the surface.
4. Rinse with water or wipe with a cloth moistened with water; Then wipe it with a dry cloth.
5. Mix a disinfectant solution compatible with the sensor according to the instructions on the disinfectant label for solution concentration and
disinfectant contact time.
6. Clean the surfaces with a disinfectant solution.
7. Air dry.
8. Check the sensor and cable for damage such as cracks, splits, or fluid leakage. If damage is detected, stop using the sensor and contact
SonoSite or your local representative.
Cleaning and disinfecting the transducer (immersion method)
1. Disconnect the drive from the system.
2. Remove all parts of the sensor housing.
3. Wipe the surface with a soft cloth lightly dampened with mild soap or a compatible cleaning solution to remove any particles or body fluids.
Apply the solution to the cloth, not the surface.
4. Rinse with water or wipe with a cloth moistened with water and then wipe with a dry cloth.
5. Mix a disinfectant solution compatible with the sensor according to the instructions on the disinfectant label for solution concentration and
disinfectant contact time.
6. Immerse the sensor in the disinfectant solution no more than 12-18 inches (31-46 cm) from where the cable enters the connector. Follow the
instructions on the disinfectant label when immersing the probe.
7. Rinse until submerged as directed on the disinfectant label, then air dry or dry with a clean cloth.
8. Check the sensor and cable for damage such as cracks, splits, or fluid leakage. If damage is detected, stop using the sensor and contact
SonoSite or your local representative.
Thank you so much for your attention .If you're interested in SonoSite HFL38X 13-6 MHZ Ultrasound Transducer REF P07682-20A ,feel free to contact us .Thank you . |
||||||||||||||||
| Product Tags: 6 MHZ Ultrasound Transducer REF P07682-20A Ultrasound Transducer |
|
CHISON D3C60L Ultrasound Convex Array Transducer Probe for Q Series System Machine Used with Good Condition |
|
Used Good Condition Toshiba PLM-703AT Linear Array Ultrasound Transducer Probe 7.5 Mhz Suitable For Emergency Rooms |
|
PHiliph C9-2 Probe Broadband Curved Array Transducer for Affiniti 30 Ultrasound Machine System 2 - 9 MHz |
|
Samsung Medison EC4-9IS Endocavity Ultrasound Transducer Probe for Accuvix A30 2013 |
|
GE M5S-D Sector Array Probe Matrix Phased Array Transducer for Voluson E8 Vivid E9 Logiq E9/S8 1.0 - 5.0 MHz |
|
SonoScape 3C-A Convex Probe Curved Array Transducer for Ultrasound Machine Abdominal Obstetrics Gynecology |