P/N 930 117 17 Defibrillator Machine Parts LCD Display Assembly
|
|
GE Marquette CardioServ LCD Display Assembly P/N 930 117 17
Signal display Backlit LCD display, 2-channel clear mode, left corner calibration pulse (for ECG) Alphanumeric display of alarm messages, sensitivity, leads, twitch flashes, alarm limits, heart rate, energy and softkey labels Freeze ECG frame, simultaneous display of ECG current and (SpO 2- for compatible devices) smaller scale plethysmographic ECG trace, 1.5 times larger than recorded : sensitivity 1 cm/mV, signal amplitude 1 mV on screen 1.5 cm * Eraser strip scanning speed 25 mm/s * Trace length is 4.6s in real-time mode * Screen size: 115mm wide, 86mm high * 320 x 240 pixel resolution (0.36 x 0.36 mm spacing) * The displayed image can be rotated 180°
LCD without display - Wrong contrast setting? (adjust contrast) - Is the contrast button defective? (to do a key test) - Are the LEDs or data ports OK? - Faulty lighting? - Fuse SI500 (16V), SI501 (backlight), SI502 (5V) on the Digital board OK? - Is the screen broken? - Wrong PCB numbers?
To remove the display: - Unplug the power cord to expose the light. - Carefully pull up the safety lock of the de-energized plug on the display membrane keyboard plug, then remove the cable from the membrane keyboard. - Remove the 4 screws from the screen
Specification
Testing display accuracy of SpO2 and pulse rate Check displayed values on the LCD screen using the following simulator settings: Simulator value SpO2 Value displayed 95.5% 92% - 99% 90.6% 87% - 94% 85.5% 82% - 89% Simulator value pulse rate Value displayed 70 bpm 66 bpm - 74 bpm 100 bpm 96 bpm - 104 bpm 160 bpm 155 bpm - 165 bpm
Activate the interference test on the simulator by pressing and holding the button, the SpO2 value should remain on the screen. Remove the SpO2 plug from the monitor. A “No Sensor” message is displayed with an intermittent alarm tone when the alarm is activated. Press the (F4) button, the alarm will turn off. Connect the digital probe and check if it works properly.
RFQ
Our MOQ is one piece .Welcome to take samples to test our quality .
Usually it will take around 2-3 working days to prepare for the order after payment and you will receive it around 5-7 working days .
Sure .We can ship to your Chinese forwarder free .
Sure .Video can be sent as well before the shipment
If you're interested in GE Marquette CardioServ LCD Display Assembly P/N 930 117 17,feel free to contact us .Thank you .
|
||||||||
| Product Tags: P/N 930 117 17 Defibrillator Machine Parts LCD Display Assembly Defibrillator Machine Parts |
|
Innomed Cardio-Aid 200-B Defibrillator Energy Release Board R-4721-6 M50812003 with Good Condition |
|
GE DASH2000 monitor complete set of housings patient monitor complete cover medical part |
|
Original Medtronic LP20 Defibrillator Paddle Pad with 2-360 J Energy Range 8mm Thickness and ≥95% Efficiency |
|
Omron AED Cover With Key Membrane And Speaker W.321414.UK C9457 QC PASSED HQ17-013-007 |
|
Medtronic LP20 Display Screen Protector In Good Working Spare Parts Medical |
|
Innomed Cardio-Aid 200-B Defibrillator Test Board ARTEMA DEFI TEST 39140-B Used in Good |