MSDS Novel Coronavirus Detection Kit
|
|
Novel Coronavirus (COVID-19) Nucleic Acid Detection kit
Novel Coronavirus (COVID-19) Nucleic Acid Detection Kit is designed for qualitative detecting the RNA of SARS-CoV-2 in the samples of nasal/throat swab.
For suspected cases, suspected cluster cases, or other cases that require the diagnostic and differential diagnosis of SARS-CoV-2 infection, which helps the clinical diagnosis of the infection of SARS-CoV-2.
【PACKING】
50 tests/kit & 200 tests/kit
【INTENDED USE】
This kit is used for the detection of COVID-19 (ORF1ab/N Gene) Nucleic Acid and assisted diagnosis and epidemiological surveillance of COVID-19.
【TEST PRINCIPLE】
The kit adopts the principle of real-time fluorescent quantitative PCR technology, designs specific primers and probes for COVID-19 (ORF1ab/N gene), and combines them with real-time fluorescent quantitative PCR instrument to detect the nucleic acid of COVID-19 virus. So as to realize the qualitative detection of COVID-19 virus nucleic acid. In addition, the PCR detection system uses the positive internal control, which monitors the presence of PCR inhibitors in test specimens by detecting whether the internal control signal is normal, to avoid a false negative result.
【COMPONMENTS OF THE DIGNOSTIC KIT】
1. Do not mix or exchange components from different kit lots. 2. The CODID-19 Negative Control is sterile nuclease-free water, and the COVID-19 Positive Control is transcribed RNA in vitro which contains target genes ORF1ab, N gene, and Internal Control gene.
【STORAGE AND STABILITY】
1. The diagnostic kit should be stored in a sealed pouch at -20 ± 5℃. The expiry date is 12 months. 2. Please refer to the date of manufacture and expiry date on the outer package. 3. The reagent keep valid and stable within the expiry date if not used. The kit should not be frozen and thawed more than 5 times.
【APPLICABLE INSTRUTMENT】
The diagnostic kit is applicable to MA-6000, ABI series, Bio-Rad series, Roche Light Cycler R480, Cepheid Smart Cycler, Rotor-Gene series and other multi-channel real-time quantitative PCR instruments.
【SPECIMEN REQUIREMENTS】
1. Sample types: Upper respiratory tract specimens (including throat swabs, nasal swabs, nasopharyngeal extracts, deep cough sputum), lower respiratory tract specimens (including respiratory tract extracts, bronchial lavage fluid, alveolar lavage fluid, lung tissue biopsy specimens), Tissue culture and other samples.
2. Storage conditions: The collected specimens should be submitted for inspection in a timely manner, and the specimens should be stored at 4°C within 24 hours. It is best to store at -70°C for more than 24 hours, and avoid repeated freeze-thaw cycles.
【TEST METHOD】
1. 1 Take out each components from the diagnostic kit and place them at room temperature. Allow the reagents to equilibrate at room temperature, then vortex each of them respectively for later use. 1.2 According to the quantity of test specimens, COVID-19 Positive Control and COVID-19 Negative Control, pipette appropriate quantity of COVID-19 PCR Mix and COVID-19 Enzyme Mix ( COVID-19 PCR Mix 15 μl/test+ COVID-19 Enzyme Mix 5 μl/test), mix them thoroughly to make a PCR-Master mix, centrifuge it instantaneously for later use.
1.3 Transfer the above-prepared reagents to the “specimen processing region” for later use.
2. Processing and loading of specimens (performed at “specimen processing region”)
2.1 This diagnostic kit does not include Viral RNA&DNA Extraction Kit. It is recommended to use Viral RNA&DNA Extraction Kit to extract viral RNA. The specific operation is in accordance with its instructions.
2.2 Add 20 μl PCR-Master mix into PCR reaction tube with 5 μl above processed sample, COVID-19 Positive Control and COVID-19 Negative Control, and cap the tube. Carry out fluorescence quantitative PCR detection on fluorescence PCR instrument.
3. PCR Amplification (performed at “nucleic acid amplification area”)
3.1 Place PCR reaction tubes into the specimen wells of the amplification equipment. Setup COVID-19 Positive Control, COVID-19 Negative Control and specimens to be tested in the corresponding sequence and input specimen name.
3.2 Set cycle parameters according to the following table for PCR amplification.
Note:
1)The fluorescence collection is set at "Step 3: 55 ° C, 30s". Selection of detection channels: FAM, HEX and Cy5, where FAM channel is ORF1abgeneand HEX channel is N gene. Cy5 channel is the internal control gene, and the reaction system is set to 25 μl. ABI series fluorescent PCR instruments do not select ROX calibration and select None for the quenching group.
1. Condition Setting for Result Analysis
The adjustment principle of Baseline and Threshold is generally based on the results of the automatic analysis of the instrument. When the overall slope of the curve appears, the Start, End, and Threshold values of the Baseline can be adjusted according to the image. Usually, the user can adjust it according to the actual situation. The Start value can be set to 3 - 15, and the End value can be set to 5 - 20. Adjust the amplification curve of the negative control to make it straight or below the threshold line.
2. The Validity of The Kit
2.1 The COVID-19 Positive Control: FAM, HEX and Cy5 channels have typical S-type amplification curves and Ct≤32. 2.2 The COVID-19 Negative Control: FAM, HEX and Cy5 channels have no Ct or Ct >38. 2.3 Note: The above conditions must be met at the same time, otherwise this experiment is invalid and needs to be repeated.
3. The Positive Judgment Value Through the study of reference values, it was determined that the Ct reference value of the target gene and the internal control gene detected by this kit was both 38.
The Sample Result Judgment
1. If the test sample detects a typical S-type amplification curve in the FAM, HEX and Cy5 channels and the Ct is ≤38, the sample can be judged to be COVID-19 positive. 2. If the test sample has no Ct or Ct > 38 in the FAM and HEX channels, and there is a typical S-type amplification curve in the Internal Control channel (Cy5), Ct≤38, the sample can be judged to be COVID-19 negative. 3. If the test sample only has a Ct value ≤38 in a single channel of the FAM or HEX channel, and there is a typical S-type amplification curve in the Internal Control channel (Cy5), Ct ≤38, the results need to retest. If the results repeated are consistent, the sample can be judged as COVID-19 positive, the results repeated are negative except for the typical S-type amplification curve of the Internal Control channel(Cy5), Ct≤38, which could be judged as COVID-19 negative. 4. If no typical S-type amplification curve (No Ct value) or Ct value> 38 is detected in the FAM, HEX, and Cy5 channels of the test sample, it means that there is a problem with the quality of the sample or a problem with the operation. If the result is invalid, you should find and eliminate the cause, collect the sample again, and repeat the test (if the test result is still invalid, please contact the company).
【Limitations of Detection Method】
1. Test results of the diagnostic kit can be used only for clinic reference. The clinical diagnosis and treatment of patients should be considered in conjunction with their symptoms, signs, medical history and other related conditions. 2. False negative results may occur when the concentration of the detected nucleic acid in the test sample is below the minimum detection limit of this kit. 3. Improper handling of the tested sample during collection, transportation, storage, and processing can easily result in RNA degradation and false negative results. 4. When samples are cross-contaminated during collection, transportation, storage, and processing, it is easy to get false positive results.
【Product Performance Index】
1. LOD: The limit of detection is 200 copies/ml. 2. Precision: Coefficient of variation(CV%) of precision Ctvaluewithinbatch≤3%. 3. Specificity: There is no cross reaction between the kit and positive samples, such as Human Coronavirus HCoV- NL63, Human Coronavirus HCoV-OC43, SARS Coronavirus, MERS Coronavirus, Influenza A Virus, Influenza B Virus Yamagata, Victoria, H1N1 Influenza Virus, H3N2 Influenza Virus, H5N1 Influenza Virus, H7N9 Influenza Virus, Respiratory Syncytial Virus A, Adenovirus(type 2), Adenovirus (type 2), Mycoplasma Pneumoniae, Chlamydia Pneumoniae, Pertussis, Streptococcus Pneumoniae, Rhinovirus(Type A) etc.
【PRECAUTIONS】
1. The entire detection process should be performed strictly in accordance with the requirements of this manual in the reagent preparation area, sample processing area, and PCR amplification area, and the experimental clothes, instruments, and consumables in each area should be used independently and cannot be mixed. 2. To avoid RNA degradation, the sample processing process should be operated at 0-4℃, and the test should be performed immediately after the experiment is completed. Utensil consumables used in sample processing should be nuclease-free. 3. Negative and positive controls should be set for each experiment. 4. All reagents in the kit should be fully thawed and mixed at room temperature and centrifuged immediately before use. 5. All negative and positive controls in the kit should be transferred to the sample preparation area and stored separately before the first use. 6. To prevent fluorescence interference, avoid touching the PCR reaction tube directly with bare hands, and avoid any markingonthe PCRreactiontube. 7. The instrument amplification related parameters should be set in accordance with the relevant requirements of this manual, and different batches of reagents cannot be mixed. 8. The product waste during the experiment should be detoxified before being discarded.
【INDEX OF SYMBOL】
【EXPORTER】 Magnus Internationa Limited
F12, New City Internationa Mansion A, 234 Huapao Ave. Liuyang, Hunan Province 410300 China
Contact: Goodwellmedical@gmail.com
【AUTHORIZED REPRESENTATIVE】
Lotus NL B.V. Koningin Julianaplein 10, 1e Verd, 2595AA, The Hague, Netherlands. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Product Tags: MSDS Novel Coronavirus Detection Kit Renji Novel Coronavirus Detection Kit | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
MSDS Novel Coronavirus Detection Kit |
|
One step PCR Nucleic Acid Testing Kit With Pharyngeal Specimens |
|