Portable Multi Model Better Price High Quality RBH-800 Syringe Pump Technology Parameters of Syringe Pump
Main features | Reliable and Safe: | Advanced safety design able to stop infusion and alarm even if the
CPU damage,the medical staff can use it safely. | Applicable syringe | 5ml,10ml,20ml,30ml,50/60ml | The maximum flow rate range | 5ml syringe: 0.1ml/h~200ml/h | 10ml syringe: 1ml/h~300ml/h | 20ml syringe: 1ml/h~400ml/h | 30ml syringe: 1ml/h~600ml/h | 50/60ml syringe: 1ml/h~1200ml/h | Precision | Better than ±2%; | (but 5ml syringe and <1ml/h: Better than ±8%) |
Parameter range | 1, Flow-rate: 0.1~1200.0 ml/h, 0.1ml/h step; | 2, Dose: 0.1~999.9 mg, 0.1mg step; | 3, mg/kg/h: 0.001~999.999 mg/kg/h, 0.001 mg/kg/h step; | 4, Body-weight: 1~300 kg, 1kg step; | 5, KVO rate: 0.1~5.0 ml/h, 0.1 ml/h step; | 6, Bolus rate: 0.1~1200.0 ml/h, 0.1 ml/h step; | 7, Bolus volume: 0.1~99.9 ml/h, 0.1ml step | Alarm | 1, Occlusion,0~175KPa, 8 grades of sensitivity; | 2, Pressure sensor is abnormal; | 3, Push rod fall off; | 4, The drug running out; | 5, The injection will finish; | 6, The motor rotation fault; | 7, Overtime pause; | 8, Low battery; | 9, Battery failure; | 10, Battery temperature is too high; | 11, Mains missing tips; | 12, MCU fault, and stop injection. | Safety classification | Class I; Internal power supply, CF type, IPX4 equipment |
Power supply | AC 100~240V, 50/60Hz, Built-in rechargeable Li-battery | Power consumption:20VA | Working time of battery: 10h@5ml/h | Weight and size | 37cm(w)*24cm(h)*21cm(t); about 2.1kg | Optional accessories | Wireless communicating module. | Chinese Certificate | yes |
|