X-ray tube | Target: Cu,Fe,Co,Cr,Mo,W (Metal or Ceramic optional ) |
Focus size: 1*10mm2 or 0.4*14mm2 |
Rated power:2.4kw or 2.7 kw |
X-ray generator (PLC Control) | Maximum output power: 4kw high frequency high voltage generator |
X-ray tube voltage:10~60kv, 0.1kv/step |
X -ray tube current: 5~80mA, 0.1mA/step |
Output stability ≦0.05% (mains fluctuation ±10%) |
Goniometer | Goniometer structure: vertical goniometer θs-θd (sample fixed) scanning radius: Standard 225mm (185- 300mm continuously adjustable |
Scanning range: -35°~160° |
Scan speed: 0.0012°~120°/min |
Positioning speed: 1500°/min |
Drive mode: θs-θd linkage, θs or θd single |
Scan mode: Stepping, continuous, segmentation, transmission
scanning |
Minimum step angle: 0.0001° |
measurement accuracy: ≦0.001° |
20 angle repeatability: 0.0001° |
Record control unit (detector) | Detector Type: SDD or SC |
Maximum linear count rate: 1X107cps,( noise≤0.2cps) |
Energy resolution: SC ≤50%,PC ≤2 |
Working mode: PLC automatic control differential, integral mode
conversion |
PLC automatically performs PHA dead time correction |
Detector high voltage 0~2200VDC continuously adjustable |
Voltage stability: 0.005% |
Safety protection alarm type | KV and mA are too high or too low, overload protection, no water
alarm, X-ray tube over-temperature and other multiple protection,
to prevent radiation leakage, protective door interlock protection
and platform lead shielding protection. |
X-ray leakage stability Dimensions (main body | ≦0.1 µSv/h (maximum power of X-ray tube, without subtracting the
background) |
≦0.3% (40kV, 40mA, continuous operation for 8 hours) |
1300mmx1000mmx1800mm (length * width* height) |