| JVVMEG-HCD75-MP-RFM is equipped with air cleaning system of negative pressure exhaust
air;It is used for radiation shielding in the repacking process of
radioactive drugs. The heating chamber is composed of three parts. The control area is
located in the upper part, and the air cleaning unit and high
efficiency air filtration unit are installed. The control panel is
easy to use and observe by the operator. The working area is
located in the middle, and the front is installed with lead glass
Windows and two handling hand holes for entering the working area;
The auxiliary area was located in the lower part, and the shielding
well of the activity meter and ionization chamber and the
radioactive material delivery device were installed, which fully
met the use function of radioactive medicine in the process of
repacking. |
Main Technical Parameters |
·cGMP biological cleanliness level working area |
1) | Configured with vertical laminar negative pressure ventilation
system Material as Integrated SS314 profile |
2) | Cleanliness: Class A grade (IS0100) |
3) | Air supply model:Vertical laminar flow |
4) | Laminar wind speed: 0.35-0.56m /s |
5) | Demand air volume :500 m3/h |
6) | Sealing way:Air inflating |
7) | Exhaust air filtration: fully efficient HEPA99.995% @0.5um (H14) |
8) | Exhaust air filtration:Radioactive gas filter adsorption |
·Lead shielding conforming to ICRP radiation environmental safety
standards |
1) | Front side:75mmPB |
2) | Left/right side:70mmPB |
3) | Target line shielding:60mmPB |
·Ergonomically designed operation control panel |
1) | Control: The human-machine interaction interface controls and feeds
back all information |
2) | Visible lead glass Windows; 320 mm(w)*220 mm(h) |
3) | Dose monitoring display: integrated in human-machine interaction
interface |
·Size |
1) | Internal dimensions :850 mm (w)*590 mm (d)*680 mm (h) |
2) | External dimensions: 1125 (w)*1078(d) *2400 (h) |
3) | Accessories: lamp, UV lamp, ozone sterilization device, power
socket, etc |
|
Cutting-edge Technology Introduction |
·Protection Delivery Device of Radioactive Medicine |
1) | Function: No need to open the door of the hot cell, the radioactive
materials or lead cans can be transferred into the working area (or
removed from the working area) in a low state,so that reducing the
intensity of manual operation and unnecessary radiation, and
improving the efficiency of dispensing. |
2) | Shield tray: stainless steel 304, Dia115 mm, 60mm pb |
3) | Transmission mode: pneumatic/linear transmission |
·Radiation Protection of Ionization Chamber of Active
Meter=Radioactive protective Well |
1) | At the bottom of radioactive well, there is a lifting adjustment platform to fix or adjust the height of the ionization chamber of the activity meter |
2) | Radioactive protective well: 70mmpb |
·Real-time dose monitoring and radiation safety interlock device |
1) | Function: Real-time display of thermal room radiation level,
provide radiation warning. In addition, the dose threshold can be
preset and the shielding door of the thermal chamber can be
interlocked to prevent unnecessary radiation to the personnel and
the environment caused by incorrect operation |
2) | Built-in 1 semiconductor detector |
3) | Energy range: 50 KeV-- 1.5 MeV |
4) | The dose range: 0 uSv/h to 9999 uSv/h |
5) | Threshold setting: Arbitrarily set the linkage threshold |
Parameters Of Manipulators The joint manipulator has 7 independent movement degrees of freedom
in space (X-direction movement, Y-direction movement, Z-direction
movement, azimuth rotation, clamp pitch, clamp rotation, clamp
opening/closing). The operation is light and flexible, and the balance device of the
manipulator meets the requirements of the balance of the operation
load. |