- Accuracy
- Standard: ± 0.08 in. (± 2 mm)
Ultra accuracy: ± 0.04in. (± 1mm)
- Measurement Range
- 492 ft (150m), 82 ft (25 m) for Rosemount 5408:SIS in Safety mode
- Operating Pressure
- -15 to 1450 psi (-1 to 100 bar)
- Operating Temperature
- -76 to +482 °F (-60 to +250 °C)
- Power Supply
- 4-20mA/HART: 12-42.4 VDC (12-30 VDC in Intrinsically Safe
installations)
Foundation™ Fieldbus: 9-32 VDC (9-30 VDC in Intrinsically Safe
installations and 9-17.5 VDC for FISCO)
- Communication Protocol
- 4-20mA HART® (2-wire loop-powered), Foundation™ Fieldbus
- Certifications/Approvals
- ATEX, IECEx, USA, Canadian, 3-A, CRN, see full specs for complete
list of certifications
- Safety
- IEC 61508 certified to SIL 2
TÜV tested and WHG approved for overfill prevention
- Antenna type
- Cone antenna, Process seal antenna, Parabolic antenna
- Diagnostics
- Enhanced diagnostics enabling proactive maintenance: Signal Quality
Metrics, Smart Meter Verification
- Warranty
- Up to 10-year limited warranty
Features - Low 12V dc lift-off voltage
- Immune to intermittent power loss
- Unique PTFE seal design eliminates the need for o-rings
- Model tailored to SIS requirements, certified to IEC 61508 SIL 2
- Solid measurement with unique solids algorithm
- Signal Quality Metrics: Proactively alerts the operator ahead of
potential issues such as antenna coating / build-up, enabled
preventative maintenance to be scheduled
- Smart Meter Verification: Built-in automatic diagnostic tool
ensures transmitter health and integrity without interrupting the
process. Provides a simple pass/fail report helping identify and
resolve potential issues. May be run on a user-configurable
schedule or on demand.
- Smart Proof test package: Quick and easy periodic proof-testing
with built-in wizards, offering step-by-step guidance, dynamic
visualizations, and ability to save or print detailed reports.
- Process seal antenna for applications with heavy condensation and
aggressive medias. Also available with Tri Clamp connection.
- 3-A certified for hygienic applications
The Rosemount™ 5408 Non-Contacting Radar Level Transmitter uses
enhanced technology and Human Centered Design to deliver accurate,
reliable measurements on both liquid and solid materials. Using
two-wire FMCW technology the Rosemount 5408 deploys a continuous
echo to maximize radar signal strength and produce a more robust
and reliable measurement. Delivering ease-of-use at every touch
point, the Rosemount 5408 uses an intuitive software interface to
guide the operator through installation, commissioning,
proof-testing and maintenance. Powerful onboard diagnostics suite
proactively monitors transmitter health and alerts if there is an
issue. Safety certified to IEC 61508 for SIL 2 applications, the
Rosemount 5408:SIS reduces cost of risk, increases efficiency and
protects people and the environment.
|