DM-2000 0.25'' WC Digital Magnehelic Gauge Differential Pressure Flow
|
Dwyer series DM-1000 magnehelic differential pressure gauge 0-0.25''w.c., stainless steel
Introduction The DigiMag Series DM-1000 Digital Differential Pressure and Flow Gages monitor the pressure of air and compatible gases just as its famous analog predecessor the Magnehelic Differential Pressure Gage. All models are factory calibrated to specific ranges. The 4-digit LCD can display readings in common English and metric units so conversions are not necessary. The simplified four button operation reduces set up time and simplifies calibration with its digital push button zero and span. The DigiMag Digital Gages are more versatile than analog gages with their ability to be field-programmed to select pressure, air velocity or flow operation depending on model. The DigiMag Digital Gages have an added feature for filter applications where a set point can be input where the display will blink when the filter is dirty, alerting the user that a maintenance action needs to occur.
Specification
Working principle The Series DM-1000 DigiMag® Digital Differential Pressure and Air
Flow Gages possess a full-scale accuracy of 1% on ranges down to 2
in w.c. and 2% accuracy down to the very low ranges of 1 in to 0.25
in w.c. DigiMag® Digital Differential Pressure Gages offer power
versatility by working with 9-24 VDC line power or simply 9V
battery power. If using line power and connecting the 9V battery,
the battery will act as a back-up if line power is lost or
interrupted.
The DigiMag® Digital Gages are more versatile than analog gages
with their ability to be field-programmed to select pressure, air
velocity or flow operation depending on model. The DigiMag® Digital
Gages have an added feature for filter applications where a set
point can be input where the display will blink when the filter is
dirty, alerting the user that a maintenance action needs to occur.
Performance characteristic
Service: Air and non-combustible, compatible gases. Wetted Materials: Consult factory. Housing Materials: Glass filled plastic. Accuracy: ±1% FS including linearity, hysteresis and repeatability; ±2% FS for ranges 1 in w.c. and below. Temperature Limits: 0 to 140°F (-18 to 60°C). Compensated Temperature Limits: 32 to 122°F (0 to 50°C). Long Term Stability: ±1% FS per year. Thermal Effect: ±0.05% FS/°F typ.; ±0.10% FS/°F for ranges 1 in w.c. and below. Display: 4-digit LCD (digits: 0.60H x 0.33W). Display Update: Selectable for 1 second to 10 minutes or update only from button push. Pressure Limits: Normal and bi-directional ranges 5 in w.c. and lower = 2 psi (13.7 kPa); Normal and bi-directional ranges 10 in w.c. and higher = 11 psi (75 KPa). Selectable Engineering Units: in w.c., psi, kPa, Pa, mm w.c., mBar, in Hg, mm Hg, FS (0-100%). Power Requirements: 9 V alkaline battery, included, user replaceable or external power supply 9-24 VDC. Battery Service Life: Battery life depending on the display update setting: 150 hours (typical) if display update = 1 second; 9 month (typical) if display update = 10 minutes; 1.5 years (typical) if display update is disabled. Battery may last up to four times longer when using lithium-based battery ULTRALIFE U9VL-J. Current Consumption: 5 mA max. Electrical Connections: Removable terminal block for 16 to 26 AWG. Electrical Entry: Cable gland for 0.114 to 0.250˝ (2.9 to 6.4 mm) diameter cable. Process Connections: 1/8˝ (3 mm) ID tubing. Enclosure Rating: NEMA 4X (IP66). Weight: 1.18 lb (535 g). Size: 5˝ (127 mm) OD front face. Agency Approvals: CE.
Application areas: • Filter monitoring air flow rate • Blower vacuum detection • fan pressure indicator • Pipe room or building pressure • clean room positive pressure indicator
|
||||||||||||||||||||||||||||||
Product Tags: Flow Digital Magnehelic Gauge Differential Pressure Digital Magnehelic Gauge 0.25'' WC Digital Magnehelic Pressure Gauge |
![]() |
Barksdale 0417-011 Pressure Switch 8111-PL1-B With Good Price |
![]() |
Accurate Oxygen Detection Cosmos Oxygen Indicator XP-3180 With Extractive Sampling Method |
![]() |
Fluke 376 FC True-RMS Clamp Meter with iFlex |
![]() |
Endress+Hauser E+H CYK10-A051 PH Digital Electrode Cable Digital Measuring Cable 5 m |
![]() |
Mettler Toledo InPro4800i/SG/120 52005383 PH Sensor |
![]() |
GF Signet 3-2725-60 2724-2726 pH and ORP Electrodes |