1000L High And Low Temperature And Humidity Chamber -70℃

Brand Name Sinuo
Model Number SN881-1000L
Certification Calibration Certificate (Cost Additional)
Place of Origin China
Minimum Order Quantity 1
Payment Terms T/T
Supply Ability 10 Sets Per Month
Delivery Time 30 Days
Packaging Details Export packing
Humidity fluctuation ≤2.5%
Temperature fluctuation ±0.5℃
Cooling Rate ≧1°C/min
Heating Rate ≧2-3°C/min
Humidity Range 20%~98%RH
Temperature Range -70 °C ~+150°C
Inner Size 1000*1000*1000 mm
External Size 1280*2150*2050 mm
Detailed Product Description

 

 

1000L High and Low Temperature and Humidity Chamber -70℃

 

Complied Standards & Clauses


 

It conforms to standards:

IEC60068-2-1:2007 “Environmental Testing-Part 2-1: Tests-Test A: Cold, IDT”

IEC60068-2-2:2007 “Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT”

IEC60068-2-78:2012 “Environmental Testing-Part 2-78: Tests-Test Cab: Damp heat, steady state, IDT”

IEC60068-2-30:2005 “Environmental Testing-Part 2-30: Tests-Test Db: Damp heat, cyclic (12h+12h cycle), IDT”

 

Product Overview


 

This test chamber is used for cold, heat, humidity and dryness testing of raw materials and devices such as electronics, electricians, electrical appliances, plastics, etc. The purpose is to confirm the products reliability of storage and transportation under high and low temperature alternating heat and humidity.

It conforms to standards:

IEC60068-2-1:2007 “Environmental Testing-Part 2-1: Tests-Test A: Cold, IDT”

IEC60068-2-2:2007 “Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT”

IEC60068-2-78:2012 “Environmental Testing-Part 2-78: Tests-Test Cab: Damp heat, steady state, IDT”

IEC60068-2-30:2005 “Environmental Testing-Part 2-30: Tests-Test Db: Damp heat, cyclic (12h+12h cycle), IDT”

 

Technical Parameters


 

Electrical control system
a) Controller modelCold balanced touch-type intelligent programmable temperature controller (power saving control)
b) Screen introduction

1. Curves and temperature and humidity can be displayed on the same interface

2. The screen display of temperature and humidity 


3. Curves can be recorded in real-time

 

c) Controller specifications

Technical indicators:

1. 7-inch color touch screen

2. Two control modes: program/fixed value

3. Sensor type: two PT100 inputs (electronic sensor input optional)

4. Output mode: voltage pulse (SSR)/control output: 2 (temperature/humidity)/2 4-20mA analog output/16 relay outputs (passive)

5. Control signal: 8 IS control signals/8 T control signals/4 AL control signals

6. Alarm signal: 16 DI external obstacle alarms

7. Temperature measurement range: -90.00℃--200.00℃, (optional -90.00℃--300.00℃) error ±0.2℃

8. Humidity measurement range: 1.0%--100%RH, error ±1%RH

9. Communication interface: (RS232/RS485, maximum communication distance 1.2km [optical fiber up to 30km])

10. Interface language type: Chinese/English

11. Equipped with English input function

12. Can be used with a printer (USB function optional)

13. A variety of signal combination relay outputs, and the signals can be subjected to logical operations (NOT, AND, OR, NOR, XOR), referred to as PLC programming ability

14. Diversified relay control methods: parameter->relay mode, relay->parameter mode, logic combination mode, composite signal mode

15. Program editing: 120 groups of programs can be programmed, and each group of programs can program up to 100 segments

Controller technical parameters

Accuracy: temperature ±0.1℃+1digit, humidity ±1%R.H+1digit

Resolution: temperature ±0.01℃, humidity ±0.1%R.H.

Temperature slope: 0.1~9.9 can be set.

It has upper and lower limit standby and alarm functions.

Temperature and humidity input signal dry and wet bulb PT100x2.

9 sets of P.I.D control parameter settings, P.I.D automatic calculation.

Dry and wet bulb with automatic correction screen function

Control method

1. Anti-integral saturation PID

2. BTC balanced temperature control method + DCC intelligent cooling capacity control + DEC intelligent electrical control (temperature test equipment)

3. BTHC balanced temperature and humidity control method + DCC intelligent cooling capacity control + DEC intelligent electrical control (temperature and humidity test equipment)

 