Nominal internal volume | 600L |
Effective size of inner box (cm) | 120×100×50(L×W×H) cm |
Outline space (cm) | 210×125×140(L×W×H) cm |
Weight | About 150kg |
Maximum power | 3KW |
Maximum current | 14A |
Power supply |
Power supply conditions and power supply | 1. AC 220V single-phase two-wire + protective grounding; allowable
voltage fluctuation range ±10%V; 2. The allowable frequency fluctuation range is 50±0.5HZ; TN-S mode
power supply or TT mode power supply. 3. The grounding resistance of the protective ground wire is less
than 4Ω 4. The user is required to configure an air or power switch of
corresponding capacity for the equipment at the installation site,
and this switch must independently control the use of the
equipment. at the installation site, and this switch must
independently control the use of the equipment. |
Performance index |
Use environmental conditions | 1. The ambient temperature is 5~30℃ and the relative humidity is
≤85%R.H; 2. The installation site must be a flat, vibration-free ground; 3. The equipment needs to be kept away from heat sources and
flammable and explosive substances; 4. The installation location must not be exposed to direct sunlight
and maintain indoor air circulation; 5. The equipment installation site needs to be clean and cannot be
installed in dusty places or dust exhaust outlets. |
Temperature and humidity range | Test room temperature: RT~50℃ Pressure barrel temperature: RT~ 63℃ Relative humidity ≥85%R.H |
Heating rate | Test room temperature RT→+50℃≤60 min Pressure barrel temperature RT→+63℃≤60 min |
Main technical parameters | Temperature resolution: 0.1℃ Temperature deviation: ±1℃ Temperature uniformity: 2℃ Temperature fluctuation: ±1℃ Spray volume (m1/80cm2/h) 1.0~2.0 (collect at least 16 hours, take
the average value) Spray pressure: 1.00±0.01kgf/cm2 Potion PH value: Neutral 6.0~7.0 Acidic 3.0~3.1 Spray solution pH value: Neutral 6.5~7.2 Acidic 3.1~3.3 Control time: 1S ~ 9999H can be set arbitrarily Note: The loss of carbon dioxide in the solution during spraying
may cause changes in pH value. You can use the following methods,
such as heating the solution to above 35°C before placing it in the
test equipment, or using fresh boiling water to prepare the
solution to reduce the carbon dioxide content in the solution. ,
can avoid changes in pH value. In order to ensure the pH value of
the spray solution in the acid test or copper accelerated salt
spray test, the pH value of the configured potion can be adjusted
to 2.8~3.0 and check whether the solution and/or solute meets the
requirements. it into the test equipment, or using fresh boiling
water to prepare the solution to reduce the carbon dioxide content
in the solution, which can avoid the change of pH value. The acid
test or copper accelerated salt spray test is to ensure the pH
value of the spray solution, Adjust the PH value of the solution to
2.8~3.0 and check whether the solution and/or solute meet the
requirements. |
Structure characteristics |
Box structure | 1. The laboratory uses imported PVC boards (Korea's Xinmeile) with
a thickness of 8mm and a durable temperature of 85°C. 2. The sealing cover of the laboratory is made of imported PVC and
acrylic window, with a thickness of 8mm. 3. The reagent refill bottle adopts a hidden water level detection
system, which is easy to clean and not easy to break. 4. The saturated air bucket is made of SUS304# stainless steel
high-pressure resistant bucket with the best insulation effect. An
external water level bar can be used to check the water level in
the saturated bucket. 5. The test material rack adopts a flat indexing rack, and the
angle can be adjusted arbitrarily to allow fogging and fogging from
all sides, allowing more specimens to be placed. 6. The sealing cover is opened and closed pneumatically. 7. The test chamber is sealed by water to ensure that corrosive
gases in the chamber do not leak. 8. The control box and test box are integrated, and the left and
right structures are easy to operate. 9. The water and electricity separation structure is adopted to
effectively prevent water from entering the electrical control box
and damaging accessories. |
Cabinet cover | The sealing cover of the laboratory is made of imported PVC board
and acrylic window, with a thickness of 8mm. The top is made into a
110° sealing cover, which effectively prevents condensation water
from dripping onto the surface of the sample during the test and
affecting the test results. |
Control panel | Youyi 7-inch touch screen controller, pressure gauge, main power
switch |
Power distribution control cabinet | Water filling solenoid valve, spray solenoid valve, demisting
solenoid valve, power distribution board, over temperature
protector, air saturation pressure barrel, spray control valve. |
Heating system | Using water vapor heating method, titanium alloy heating tube P.I.D
controls heating to achieve temperature balance. |
Water make-up system | 1. Water replenishment in the test room and pressure barrel: two
water replenishment control systems, manual and automatic, can be
connected to the tap water pipe to automatically control water
replenishment or manual water replenishment, dual replenishment
mode, easy to use; Note: If the tap water has too many impurities or is acidic or
alkaline, a filter must be installed at the water inlet. 2. Liquid replenishment: using a large hidden water tank and manual
replenishment. |
Drainage system | Drainage valve control device, manual drainage mode. |
Spray system | Using the Bernoult principle to absorb salt water and then atomize
it, the degree of atomization is uniform There is absolutely no blocking of crystallization, ensuring
continuous testing standards. Nozzle made of corrosion-resistant Teflon. |
Spray mode | Continuous/intermittent |
Flow control | The peristaltic pump can supply liquid quantitatively, and the flow
control range is: 0.1-580ML/M. |
Air supply system | The spray pressure is adjusted to 1kg/cm² in two stages. The first
time is to adjust the compressed air to 2kg/cm² and enter the
saturated barrel for heating. The second time is to adjust the
spray pressure to 1kg/cm². |
Mist exhaust system | 1. Fully automatic pressure balance mist exhaust device, just
connect the mist exhaust pipe, and the mist will be automatically
exhausted according to the pressure in the box 2. Manual defogging button switch. Just click this button before
opening the test chamber cover to quickly remove indoor fog. 3. Note: The mist exhaust pipe of the machine needs to be extended
to an open outdoor area, and it is necessary to avoid being close
to corrosive items to prevent the mist discharged from the machine
from corroding other items; the mist exhaust extension pipe must
not have wavy and discounted places, and it must be avoided inside
the mist exhaust pipe. Condensed water blocks the mist exhaust pipe
and cannot effectively discharge exhaust gas. |
Electric control system |
Controller | 1. 7-inch touch screen controller can perform fixed-value (i.e.
continuous, or constant temperature) experiments or programmable
(intermittent, cycle tests) salt spray experiments. One unit can
achieve multiple functions to meet customers' different salt spray
experiment requirements. High-precision temperature Control module,
PID automatic calculation to control temperature, high stability
platinum temperature measurement probe. 2. Standby can be set, that is, when the temperature is close to
the set temperature, the instrument automatically sprays and times
the time at the same time. Defogging will automatically start when
the time is over, and the defogging time can be set. The water
replenishment delay can effectively solve the high-frequency
replenishment actions caused by water level fluctuations. Increase
the life of the hydration system. English systems can meet the
inspection requirements of foreign customers. |
Temperature input mode | By touch screen |
Total working time | One-time cumulative timer can monitor the total working time of the
machine |
Operation mode | Continuous operation/intermittent operation. |
Temperature setting range | The upper and lower limits of the maximum temperature range are 5℃ |
Panel display | Set temperature value, current temperature value, set time, current
time, total machine working time, over-temperature abnormal
indicator light, water shortage abnormal indicator light, operation
end indicator light, spray status indicator light, defogger status
indicator light |
Temperature measuring body | PT100 platinum resistor |
Temperature control method | Anti-integral saturation PID, BTC balance temperature control
method (temperature test equipment) |
Communication method | Equipped with standard RS232 and network port communication
interfaces, and standard modbus-rtu communication protocol, it can
be realized Conduct remote communication management between central control
instruments and computers; open communication protocols and provide
communication protocols Negotiation instructions; |
Data output | 1. USB flash disk storage can be carried out through the USB
interface, and it supports U disk hot-swap and direct download to U
disk functions. function; and can copy data or directly display and print test
data/curves through a PC. 2. Data saving can be exported directly and the data will be
automatically saved; the temperature data section to be exported
can be selected through time; |
Curve record | It has battery-protected RAM that can save the device’s setting
values, sampling values and sampling time; The maximum recording time is 360 days (when the sampling period is
60S) |
Safety protection device |
Test box | Extreme over temperature protection, water shortage protection. |
Saturated barrel | Extreme over temperature protection, water shortage protection. |
Heating system | Heating pipe dry burning, abnormal water supply and drainage |
Power Supply | Leakage protection, overload, and short circuit protection |
Storage environment requirements | The environment temperature of equipment shall be kept within 0
℃~+45 ℃ |
Requirements for the installation site | The distance between the left and right side and rear side wall
panels of the test box and the wall shall not be less than 600mm,
and the distance between the front and the wall shall not be less
than 800mm. When transporting equipment, whether it can enter the
doorway, passage, elevator, etc., so as to avoid affecting your
project progress. |