- Sample limitation
| This test chamber prohibits: Testing or storage of flammable, explosive, and volatile substance
samples Testing or storage of samples of corrosive substances Testing or storage of biological samples Testing or storage of strong electromagnetic emission source
samples Testing or storage of radioactive material samples Test or storage of highly toxic substance samples Testing or storage of samples that may produce flammable,
explosive, volatile, highly toxic, corrosive and radioactive
materials during testing or storage |
- Volume, size and weight
|
- Nominal content volume
| 1500L |
- Inner chamber size
| W1000mm×D1500mm×H1000mm |
- External dimensions
| Approx. W1500mm×D2020mmxH2150mm |
- Weight
| Approx.680KG |
- Performance
|
- Temperature range
| Temperature range: -70 ℃~+180℃ |
- Humidity range
| (20 ~ 98)% RH (refer to temperature and humidity controllable range
chart, no active wet and heat load)
Temperature and humidity controllable range chart |
- Temperature fluctuation
| ±0.5℃ (The temperature fluctuation is half of the difference
between the measured maximum temperature and the minimum
temperature at the center point) |
d. Humidity fluctuation | ≤2.5% |
- Temperature deviation
| ±2.0℃ (Temperature uniformity is the arithmetic mean of the
difference between the highest and lowest temperature measured in
each test) |
- Humidity deviation
| ≤±3% |
- Temperature display resolution
| 0.01℃ |
- Humidity display resolution
| 0.1% RH |
- Heating rate
| Average 2~3℃/min (non-linear under no load) |
- Cooling rate
| Average 0.7~1℃/min (non-linear under no load) |
- Load conditions
| None |
| Sound level≤75dB (Measured in a soundproof room with an ambient temperature of 25 °
C and low echo; using A weighting, the average value of 8 points is
tested; each test point is 1 meter away from the noise source and 1
meter away from the ground) |
- Structural characteristics
|
- Material
| 1. Outer wall material: A3 plate spraying 2.Inner wall material: SUS304# stainless steel plate 3. Chamber body insulation material: 100mm high temperature
resistant rigid polyurethane foam 4. Door insulation material: 100mm high temperature resistant rigid
polyurethane foam. |
- Bottom structural strength
| The heavy capacity of the track at the bottom of the test chamber:
≤100Kg/m² (load) |
- Air conditioning channel
| ① Stainless steel long-axis centrifugal fans: 4 units/90W. ②Fan, heater, evaporator (also dehumidifier), drainage device,
pressure balance port, adjustable air guide plate, temperature
sensor |
- Standard configuration of test chamber
| ①Window 400x600x40mm 3-layer vacuum tempered glass. ②Flat embedded handle, ③Door hinge: SUS #304 imported hinge ④ Energy-saving lamps in the box: LED energy-saving lamps ⑤Lead hole: φ50mm 1 piece |
- Door
| ① The single-piece door opens outwards, with the hinge on the left
and the handle on the right (when facing the front of the box). ② Equipped with a safety door locking mechanism (the door can be
opened in the test room), the door is equipped with electric heat
to prevent condensation, and the viewing range of the insulating
glass observation window is approximately (W400×H600mm). Electric
heating device to prevent condensation on door frame |
- Control Panel
| There is a 7-inch display screen for temperature (humidity) control
and display, an operating indicator light, and a lighting button |
- Machinery room
| Refrigeration unit, compressor water tray, pressure relief device,
heating device |
- Power distribution cabinet
| Distribution board, cooling fan, main power switch, transformer,
intermediate relay, time relay, solid state relay, AC contactor,
circuit breaker |
- Heater
| ① Fin-type heat pipe-shaped stainless steel electric heater ②Heating control method: SSR (Solid State Relay) contactless equal
period pulse width modulation ③Heating power: about 4KW |
- Refrigeration system
|
- Refrigeration device
| ①American Emerson low-noise vortex ultra-low temperature
refrigeration system (air-cooled heat dissipation method). ② Cooling method: air cooling |
- Hot and cold exchange device
| Two-stage refrigeration system American Emerson low-noise scroll
compressor. Adopting ultra-high efficiency SWEP cold coal cold and
heat exchange design [made in Sweden (sus#316)], it is more
efficient than the traditional internal spiral type. |
- Heating load adjustment
| ①Adopt automatic adjustment of cold coal flow device to effectively
take away the heating load of the product to be tested and
German/Japanese technology. ②Compared with the traditional design, the control stability and
reproducibility are higher, and it achieves super efficiency of
saving power and saving more energy. |
- High efficiency components
| ①The condenser and evaporator adopt AC&R double spoiler type
aluminum fin device. ②The throttling devices adopt imported Chiming refrigeration
accessories |
- Expansion system
| Capacity controlled refrigeration system (capillary tube) |
- Evaporative condenser
| All copper tube heat exchanger |
- Refrigerator control method
| ①The controller of the control system automatically adjusts the
operating conditions of the refrigerator according to the test
conditions. ②Evaporation pressure regulating valve. ③Compressor return air cooling circuit ④The refrigeration system is designed with vibration and noise
reduction measures such as shock absorber tubes. |
- Refrigerant
| Honeywell R404a (ozone depletion index is 0) |
- Electrical control system
|
- Controller
| 7 inch smart LCD touch programmable controller |
- Controller specifications
| 1. 7-inch true color touch screen 2. Two control methods: program/set value 3. Sensor type: two PT100 inputs (electronic sensor input optional) 4. Output mode: voltage pulse (SSR)/control output: 2 channels
(temperature/humidity)/2 channels 4-20mA analog output/16 channels
relay output (passive) 5. Control signals: 8 channels IS control signals/8 channels T
control signals/4 channels AL control signals 6. Alarm signal: 16 DI external obstacle alarms 7. Temperature measurement range: -90.00℃--200.00℃, (optional
-90.00℃--300.00℃) error ±0.2℃ 8. Humidity measurement range: 1.0%--100%RH, error ±1%RH 9. Communication interface: (RS232 or RS485, the longest
communication distance is 1.2km [optical fiber can reach 30km]) 10. Interface language type: English 11. Multiple signal combination relay outputs, the signals can be
subjected to logical operations (NOT, AND, OR, NOR, XOR), referred
to as PLC programming ability; 12. Diversified relay control methods: parameter->relay mode,
relay->parameter mode, logic combination mode, composite signal
mode 13. Program editing: 120 groups of programs can be programmed, and
each group of programs can have a maximum of 100 segments |
- Controller technical parameters
| Accuracy: temperature ±0.1℃+1digit, humidity ±1%R.H+1digit Resolution: temperature ±0.01℃, humidity ±0.1%R.H. Temperature slope: 0.1~9.9 can be set. It has upper and lower limit standby and alarm functions. Temperature and humidity input signal dry and wet bulb PT100x2. 9 sets of P.I.D control parameter settings, P.I.D automatic
calculation. Dry and wet bulb automatic correction screen |
- Control method
| ①Anti-integral saturation PID ②BTC balanced temperature control method + DCC intelligent cooling
capacity control + DEC intelligent electrical control (temperature
test equipment) ③BTHC balanced temperature and humidity control method + DCC
intelligent cooling capacity control + DEC intelligent electrical
control (temperature and humidity test equipment) |
- Screen display function
| It adopts screen conversation style, no key input is required, and
the options can be directly touched on the screen. Temperature set (SV) and actual (PV) values are directly displayed. It can display the currently executing program number, segment,
remaining time and number of cycles. Operation accumulated time function. The temperature program setting value is displayed in a graphic
curve, with the function of real-time display of program curve
execution. It has a separate program editing screen, and at least 5 segments
of temperature, humidity and time can be input on each page. The fault prompt screen is displayed. The screen backlight can be adjusted. The screen display protection function can be set to timer, TIMER
or manual shutdown. |
- Program capacity and control functions
| Usable program groups: maximum 120 PATTEN. Usable memory capacity: 12000 SEGMENTS in total. Repeatable execution of commands: each command can be executed up
to 3200 times. The program is created in a conversational style, with functions
such as editing, clearing, and inserting. SEGMENTS time setting 0~99Hour59Min. Programmable timing control module device x2 set. It has a power-off program memory, and automatically starts and
resumes program execution after power is restored. With RS-485 or RS-232 communication interface. Graphical curves can be displayed in real time when the program is
executed. It has the function of automatically adjusting the freezing
capacity. It has scheduled start-up and shutdown functions. With date and time adjustment function. Button and screen lock (LOCK) function. |
- Safety protection device
|
- Refrigeration system
| 1. Compressor overheating protection 2. Compressor overcurrent protection 3. Compressor high and low pressure protection |
- Test chamber
| 1. Adjustable over temperature protection 2. Limit over-temperature protection of air conditioning channels 3. Fan motor overheat protection |
- Other protection
| 1.Leakage Protection 2.Heating tube no air drying protection 3.Overload and short circuit protection 4.Power outage protection 5.Low water level protection |
- Conditions of use
| The following conditions are guaranteed by the user |
- The installation site
| 1. The ground is level and well ventilated 2.No strong vibration around the equipment 3.No strong electromagnetic field around the equipment 4.No flammable, explosive, corrosive substances and dust around the
equipment 5. Appropriate use and maintenance space is left around the
equipment, as shown below:
A:≥40cm B:≥150cm C:≥100cm D: ≥60cm E:≥40com |
- Environmental conditions
| 1. Temperature: 5 ℃ ~ 35 ℃ 2. Relative humidity: ≤85% 3. Air pressure: 86kPa ~ 106kPa |
10.3. Power supply conditions | Power supply requirements: 1. One five-core (three-phase four-wire + protective ground wire)
cable (about 4 meters long) 2. Voltage range: Three phase AC 380V 3. Allowable frequency fluctuation range: (50 ± 0.5) Hz 4. The protective earth ground resistance is less than 4Ω 5. The user is required to configure the equipment with an circuit
breaker (63A) of the corresponding capacity at the installation
site, and this switch must be independently used by the device |
10.4. Maximum current Maximum power consumption | Imax≤32A Pmax≤13KW |
10.5. Others | Opening the door of the test chamber during the test will cause
temperature and humidity fluctuations in the chamber; if the door
is opened multiple times or is left open for a long time or the
test sample emits moisture during the test, the refrigeration
system heat exchanger may freeze and does not work. |