China FFU Fan Filter Unit manufacturer
Shenzhen Zhong Jian South Environment Co., Ltd.
We are one of the first high-tech enterprise in the industry that provides technical solutions for air pollution control and treatment.
3
Home > Products > Laminar Clean Bench >

ZJNF Class II Vertical Laminar Flow Biosafety Cabinets For Cleanrooms

Browse Categories

Shenzhen Zhong Jian South Environment Co., Ltd.

City: shenzhen

Province/State:guangdong

Country/Region:china

Tel:86--13392805835

Contact Person:
Miss.Serena
View Contact Details

ZJNF Class II Vertical Laminar Flow Biosafety Cabinets For Cleanrooms

Brand Name ZJNF
Model Number ZJNF-1
Certification CE
Place of Origin China
Minimum Order Quantity 1
Price negotiate
Payment Terms tt lc PayPal credit card
Delivery Time 18 days
Packaging Details plywood/wooden package
Server openresty
Status Code 403
Description The server understood the request, but is refusing to fulfill it.
Product Name 403 Forbidden
Detailed Product Description

How to Choose a Biosafety Cabinet for Cleanrooms

When selecting a biosafety cabinet for cleanroom environments, it is essential to consider key factors such as protection level, airflow parameters, filtration efficiency, and compliance with certifications. Below are the core selection criteria and product recommendations.


1. Key Performance Indicators

  • Airflow Velocity
    • Inflow Airflow‌: Protects the operator. For Class II biosafety cabinets, it should be ≥0.5 m/s (e.g., 0.53 m/s or 0.55 m/s) to prevent aerosol leakage.
    • Downward Airflow‌: Protects the sample, typically ≥0.25 m/s (e.g., 0.35 m/s) to ensure stable vertical laminar flow.
  • Filtration Efficiency
    • HEPA filters must capture ≥99.995% of 0.3 μm particles (e.g., 99.9995% or 99.999%) to block pathogens.
  • Certifications & Compliance
    • Must have medical device registration and meet standards like YY0569-2011 (China) or NSF/ANSI 49 (International).

2. Application Scenarios & Safety Levels

  • Class II Biosafety Cabinets‌: Suitable for BSL-2 operations (e.g., clinical testing, cell culture), protecting personnel, samples, and the environment.
  • Class III Biosafety Cabinets‌: Designed for BSL-3/4 high-risk pathogens (e.g., Ebola virus), featuring fully enclosed design and dual HEPA filtration.