Huge Memory Space Lcd Double Beam Instrument UV-1500
|
Detailed Product Description
UV -1500
The Company:
1. Why the reading is red and reach its maximum value 3.00Abs? Probably you use the glass cuvette in UV range. 2. Why the wavelength calibration or self checking fail? The deuterium lamp is turned off or there is something wrong with the D2 lamp, like the fuse, triode or the lamp itself. 3. Why is the reading is not as accurate as before? Do the wavelength calibration before testing or restore the device
to its default factory settings. 4. Why PC software can not be installed successfully? Close your anti-virus software on the computer and do it again. 5. Why the Abs value is negative(lower than zero)? Do the blanking or change the reference sample. |
|||||||||||||||||